NASA NTRS ยท 19770067329
Mass spectrometer beam system for applications in the stratosphere
Abstract
A special gas inlet system is required for reduction of high ambient pressures to ion source operating pressures for mass spectrometer investigations in the stratosphere for the determination of the abundance of minor constituents. A system has been designed which combines a mass spectrometer and a gas inlet system reducing pressure by differential pumping using high-speed liquid helium pumps and small orifices. The gas particles are formed into a neutral beam with a flag mechanism proceeding into the mass spectrometer ion source where they are ionized for subsequent mass separation. A laboratory model has been tested under conditions simulating stratospheric pressures. It was shown that the process is able to identify gases with abundances below 10 to the -8th. Some of the problems associated with gas analysis of trace constituents have been alleviated, and other applications of the system have been identified, such as pollution control and medicine.
Keep this discovery
Explore connections, maps & timelines
Mauersberger, K.. 1977-09-01. Mass spectrometer beam system for applications in the stratosphere. https://ntrs.nasa.gov/citations/19770067329