Radiometry using synchrotron radiation

Abstract

It is possible to use synchrotron radiation from electron synchrotrons and electron storage rings as an absolute source, especially in the VUV and soft X-ray regions where other standards are difficult to find. At the NBS, an electron storage ring (SURF-II) has been used to calibrate spectrometers and photometers utilized in solar and aeronomy research and in fusion plasma diagnostics. The radiation incident on these spectrometers can be calculated to uncertainties of 3%, and a technique to exactly determine the number of electrons orbiting in the ring is currently being developed to reduce this uncertainty. Detector calibrations between 5 and 55 nm are routinely performed at SURF-II and transfer standard detectors with 6-10% uncertainties over the range 5-254 nm are supplied.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Saloman, E. B., Ebner, S. C., Hughey, L. R.. 1981-01-01. Radiometry using synchrotron radiation. https://ntrs.nasa.gov/citations/19830032741