Response time correlations for platinum resistance thermometers
Abstract
The 'plunge method' recommended by ASTM has been used to determine the time constant of 100-ohm platinum resistance thermometers (PRT) considered for use in the National Transonic Facility. It is shown that the response time of ventilated PRT can be correlated with the reciprocal of the heat transfer coefficient in a given field. Universal correlations are established for the 100- and 1000-ohm PRT with uncertainties of 20 and 30 percent, respectively. The correlations are found to be consistent with the uncertainty involved in heat transfer correlations available in the literature and are recommended for use in flowing liquids and gases.