Abstract
A fiber array polar nephelometer of advanced design, the FAN I is capable of in-situ phase function measurements of scattered light from man-made or natural atmospheric particles. The scattered light is measured at 100 different angles throughout 360 degrees, thus providing a potential measurement of the asymmetry of irregularly shaped particles. Phase functions can be measured at 10 to 100 Hz rates and the range of measurable single particle sizes is from 5 micron m to as large as 8mm. For particles smaller than 5 micro m the ensemble average can be measured. The FAN I is microprocessor controlled and the data may be stored on floppy disk or printed out in tabular and/or graphical form. The optical head may be separated from the computer system for operation in field or adverse conditions. Examples of laboratory measured scattering phase functions obtained with the FAN I for spherical particles is given to illustrate its measurement capabilities.
Keep this discovery
Explore connections, maps & timelines
Farmer, W. M., Burlbaw, E. J., Deepak, A.. 1985-10-01. Particle phase function measurements by a new Fiber Array Nephelometer: FAN 1. https://ntrs.nasa.gov/citations/19860001989