Search NASAโŒ• Search

NASA NTRS ยท 19870053451

A parametric study of the beam refraction problems across laser anemometer windows

Abstract

The experimenter is often required to view flows through a window with a different index of refraction than either the medium being observed or the medium that the laser anemometer is immersed in. The refraction that occurs at the window surfaces may lead to undesirable changes in probe volume position or beam crossing angle and can lead to partial or complete beam uncrossing. This report describes the results of a parametric study of this problem using a ray tracing technique to predict these changes. The windows studied were a flat plate and a simple cylinder. For the flat-plate study: (1) surface thickness, (2) beam crossing angle, (3) bisecting line - surface normal angle, and (4) incoming beam plane surface orientation were varied. For the cylindrical window additional parameters were also varied: (1) probe volume immersion, (2) probe volume off-radial position, and (3) probe volume position out of the r-theta plane of the lens. A number of empirical correlations were deduced to aid the reader in determining the movement, uncrossing, and change in crossing angle for a particular situations.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Owen, Albert K.. 1986-01-01. A parametric study of the beam refraction problems across laser anemometer windows. https://ntrs.nasa.gov/citations/19870053451

Cite the original work for its findings. Save a collection to share your selection of sources.