Search NASA⌕ Search

NASA NTRS · 19890003465

Mass flow meter using the triboelectric effect for measurement in cryogenics

Abstract

The use of triboelectric charge to measure the mass flow rate of cryogens for the Space Shuttle Main Engine was investigated. Cross correlation of the triboelectric charge signals was used to determine the transit time of the cryogen between two sensor locations in a .75-in tube. The ring electrode sensors were mounted in a removable spool piece. Three spool pieces were constructed for delivery, each with a different design. One set of electronics for implementation of the cross correlation and flow calculation was constructed for delivery. Tests were made using a laboratory flow loop using liquid freon and transformer oil. The measured flow precision was 1 percent and the response was linear. The natural frequency distribution of the triboelectric signal was approximately 1/f. The sensor electrodes should have an axial length less than approximately one/tenth pipe diameter. The electrode spacing should be less than approximately one pipe diameter. Tests using liquid nitrogen demonstrated poor tribo-signal to noise ratio. Most of the noise was microphonic and common to both electrode systems. The common noise rejection facility of the correlator was successful in compensating for this noise but the signal was too small to enable reliable demonstration of the technique in liquid nitrogen.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Bernatowicz, Henry, Cunningham, Jock, Wolff, Steve. 1987-04-01. Mass flow meter using the triboelectric effect for measurement in cryogenics. https://ntrs.nasa.gov/citations/19890003465

Cite the original work for its findings. Save a collection to share your selection of sources.