Infrared detector responsivity measured simultaneously at multiple frequencies
Abstract
The responsivity (the ratio of the output electrical signal to the incident radiation power) of an IR detector is measured at several frequencies simultaneously. A computer-controlled variable frequency chopper is used. Chopper modulation of a blackbody is square-wave-like, hence measurements can be taken not only at the fundamental frequency but also at the harmonics. The technique gives excellent agreement with the classical technique of measuring the fundamental response alone at a set of modulation frequencies. With a small enough aperture and a high enough chopper speed, agreement can be within five percent out to 31 times the fundamental frequency. The technique allows the rapid and automated acquisition of the frequency response of IR detectors, which may then be screened for responsivity, noise, and noise equivalent power.