NASA NTRS ยท 19910036342
Subminiature infrared detector translation stage
Abstract
This paper describes a precision subminiature three-axis translation stage used in the GOES Sounder to provide positional adjustment of 12 cooled infrared detectors. Four separate translation stages and detectors are packaged into a detector mechanism which has an overall size of 0.850 x 1.230 x 0.600 inches. Each translation stage is capable of + or - 0.015 inch motion in the X and Y axes and +0.050/-0.025 inch motion in the Z axis with a sensitivity of 0.0002 inches. The function of the detector translation stage allows real time detector signal peaking during Sounder alignment. The translation stage operates in a cryogenic environment under a 10 to the -6th torr vacuum.
Keep this discovery
Explore connections, maps & timelines
Bell, Alan D.. 1989-01-01. Subminiature infrared detector translation stage. https://ntrs.nasa.gov/citations/19910036342
Cite the original work for its findings. Save a collection to share your selection of sources.