The cleanliness requirements of the Cryogenic Limb Array Etalon Spectrometer (CLAES) are described and the control program is assessed with particular emphasis given to the effects of particulate interference. The use of the CLAES is reviewed, and off-axis scatter from the earth is identified as an important noise source. Since the CLAES altitude of interest causes a low-altitude detection of 0.2 deg above the earth's surface, straylight controls are important in the development of the instrument. The development of off-axis rejection and the particulate-cleanliness level are evaluated in terms of those required for a certain BRDF. The control plan is described, and procedures are outlined including cleaning methods and facilities. The modeling predictions and procedural controls are found to effectively satisfy requirements for low scatter levels and out-of-field view-rejection by maximizing straylight control and allowing a margin against noise-equivalent radiance.