NASA NTRS · 19930050813
Data processing and calibration for tunable diode laser harmonic absorption spectrometers
Abstract
Data processing and calibration methods are described for tunable diode laser absorption spectrometers which produce harmonic absorption spectra as raw data for measuring gas mixing ratios down to parts-per-trillion levels at a variety of pressures. The methods, which take advantage of modern computer speed, memory, and data storage capabilities, are applicable to the detection of weakly absorbing gases in quantitative industrial monitoring, in addition to aircraft and balloon atmospheric measurements for which they were designed. Algorithms for calibration and data analysis, including rejection of erroneous spectra, variation of effective integration time, spectral alignment prior to averaging, and plotting and archiving of results, have been tested on actual stratospheric laser spectra recorded by the Aircraft Laser Infrared Absorption Spectrometer (ALIAS) spectrometer in numerous flights of NASA's ER-2 aircraft.
Keep this discovery
Explore connections, maps & timelines
May, Randy D., Webster, Christopher R.. 1993-04-01. Data processing and calibration for tunable diode laser harmonic absorption spectrometers. https://ntrs.nasa.gov/citations/19930050813