NASA NTRS · 19950026865
Data collection in IMS: It's not as easy as it looks
Abstract
Data collection in Ion Mobility Spectrometry (IMS) is not as easy an endeavor as it appears. Despite the advent of high speed personal computers and fast analog-to-digital converters (ADC's), care must be taken to ensure that reliable data are obtained in a timely fashion. This is especially true in hyphenated techniques, e.g. GC-IMS, where the amount of data increases dramatically when gas chromatography-ion mobility spectrometry (GC-IMS) data is being collected. Using the Graseby GC-IMS, with a gate repetition rate of 33 Hz, it is theoretically possible to collect 33 spectra per second. This collection rate is not realistically obtained due to a number of factors. Among these factors are inaccuracy of the timing signal from the IMS, the necessity to store the data, disk input/output limitations, disk operating system limitations, and program overhead. Taking these factors into account, we have achieved a data collection rate of 20 spectra per second. This paper will describe these problems, demonstrate the practical effects these problems present, and present methods for minimizing these effects.
Keep this discovery
Explore connections, maps & timelines
Davis, Dennis M., Maswadeh, Waleed, Shoff, Donald B., Harden, Charles S., Snyder, A. Peter. 1995-04-01. Data collection in IMS: It's not as easy as it looks. https://ntrs.nasa.gov/citations/19950026865
Cite the original work for its findings. Save a collection to share your selection of sources.