NASA NTRS ยท 19990111648
A Quantitative Evaluation of Dissolved Oxygen Instrumentation
Abstract
The implications of the presence of dissolved oxygen in water are discussed in terms of its deleterious or beneficial effects, depending on the functional consequences to those affected, e.g., the industrialist, the oceanographer, and the ecologist. The paper is devoted primarily to an examination of the performance of five commercially available dissolved oxygen meters. The design of each is briefly reviewed and ease or difficulty of use in the field described. Specifically, the evaluation program treated a number of parameters and user considerations including an initial check and trial calibration for each instrument and a discussion of the measurement methodology employed. Detailed test results are given relating to the effects of primary power variation, water-flow sensitivity, response time, relative accuracy of dissolved-oxygen readout, temperature accuracy (for those instruments which included this feature), error and repeatability, stability, pressure and other environmental effects, and test results obtained in the field. Overall instrument performance is summarized comparatively by chart.
Keep this discovery
Explore connections, maps & timelines
Pijanowski, Barbara S.. 1971-01-01. A Quantitative Evaluation of Dissolved Oxygen Instrumentation. https://ntrs.nasa.gov/citations/19990111648
Cite the original work for its findings. Save a collection to share your selection of sources.