Search NASA⌕ Search

NASA NTRS · 20060015669

Statistical Characterization of Environmental Error Sources Affecting Electronically Scanned Pressure Transducers

Abstract

Minimization of uncertainty is essential to extend the usable range of the 15-psid Electronically Scanned Pressure [ESP) transducer measurements to the low free-stream static pressures found in hypersonic wind tunnels. Statistical characterization of environmental error sources inducing much of this uncertainty requires a well defined and controlled calibration method. Employing such a controlled calibration system, several studies were conducted that provide quantitative information detailing the required controls needed to minimize environmental and human induced error sources. Results of temperature, environmental pressure, over-pressurization, and set point randomization studies for the 15-psid transducers are presented along with a comparison of two regression methods using data acquired with both 0.36-psid and 15-psid transducers. Together these results provide insight into procedural and environmental controls required for long term high-accuracy pressure measurements near 0.01 psia in the hypersonic testing environment using 15-psid ESP transducers.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Green, Del L., Walker, Eric L., Everhart, Joel L.. 2006-01-01. Statistical Characterization of Environmental Error Sources Affecting Electronically Scanned Pressure Transducers. https://ntrs.nasa.gov/citations/20060015669

Cite the original work for its findings. Save a collection to share your selection of sources.