NASA NTRS · 20120000655
A Portable Electronic Nose for Toxic Vapor Detection, Identification, and Quantification
Abstract
The Space Program and military use large quantities of hydrazine and monomethyl hydrazine as rocket propellant, which are very toxic and suspected human carcinogens. Current off-the-shelf portable instruments require 10 to 20 minutes of exposure to detect these compounds at the minimum required concentrations and are prone to false positives, making them unacceptable for many operations. In addition, post-mission analyses of grab bag air samples from the Shuttle have confirmed the occasional presence of on-board volatile organic contaminants, which also need to be monitored to ensure crew safety. A new prototype instrument based on electronic nose (e-nose) technology has demonstrated the ability to qualify (identify) and quantify many of these vapors at their minimum required concentrations, and may easily be adapted to detect many other toxic vapors. To do this, it was necessary to develop algorithms to classify unknown vapors, recognize when a vapor is not any of the vapors of interest, and estimate the concentrations of the contaminants. This paper describes the design of the portable e-nose instrument, test equipment setup, test protocols, pattern recognition algorithms, concentration estimation methods, and laboratory test results.
Keep this discovery
Explore connections, maps & timelines
Linnell, B. R., Young, R. C., Griffin, T. P., Meneghelli, B. J., Peterson, B. V., Brooks, K. B.. 2005-01-01. A Portable Electronic Nose for Toxic Vapor Detection, Identification, and Quantification. https://ntrs.nasa.gov/citations/20120000655
Cite the original work for its findings. Save a collection to share your selection of sources.