NASA NTRS · 20180007646

Compact LWIR Polarimeter for Cirrus Ice Properties

Abstract

A miniaturized long-wave InfraRed (LWIR) spectro-polarimeter is being developed as a prototype for the Compact Submm-Wave and LWIR Polarimeters (SWIRP) project. The polarimeter in development is a compact (20x20x40 cm) conical-scan instrument to measure the polarimetric radiation from ice cloud scattering at mm-submm (220 and 680 GHz) and IR (8.6, 11, and 12 m) bands. The LWIR polarimeter will provide a series of polarization measurements across the 8.5 - 12.5 micron band, measuring the full set of linear Stokes parameters (I, Q, U) as a function of wavelength. The spectro-polarimeter uses a combination of birefringent crystals, a Wollaston prism, a diraction grating, and an uncooled microbolometer array to measure both the degree and angle of linear polarization across the spectral bandwidth by modulating the polarization ux in wavelength with a high order retarder.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Hart, Kira A., Chipman, Russel A., Wu, Dong L.. 2018-05-14. Compact LWIR Polarimeter for Cirrus Ice Properties. https://ntrs.nasa.gov/citations/20180007646

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

First Results From an Uncooled LWIR Polarimeter for Cubesat Deployment

A compact long-wave infrared (LWIR) channeled spectro-polarimeter (IRCSP) has been developed for integration into the NASA Earth Science Technology Office (ESTO) funded submm-wave and LWIR polarimeters project to measure the microphysical properties of cloud ice. Once deployed, the IRCSP will produce the first linear Stokes measurements ( S0 , S1 , S2 ) of upper-tropospheric cirrus clouds from 8.5 to 12.5 μm. For the first time, a compact, light-weight, and uncooled LWIR polarimeter with off-the-shelf thermal optical components is demonstrated. We report narrowband calibration measurements which quantify metrics of polarimetric system performance. The response of the system to linearly polarized light is shown to agree with both a Mueller matrix model and modulation function for narrowband calibration measurements with an R2 > 0.98 from 8 to 11 μm. The polarimetric efficiency is >0.8 from 8 to 11 μm for narrowband measurements indicating satisfactory performance of the polarization optics. Beyond 11 μm, the agreement is significantly reduced as thermal noise compounds with reduced detector response. Ultimately, the observed system performance is limited by the spectral response of the detector past 11 μm in addition to the thermal noise inherent for the measurements at room temperature.

LWIR↗

Linear Stokes Measurement of Thermal Targets using Compact LWIR Spectropolarimeter

The trade-o↵ between spectral resolution and instrument throughput is analyzed for a compact, uncooled, longwave infrared (LWIR) channeled spectropolarimeter (IRCSP). The IRCSP was developed as a part of the Submm Wave and InfraRed Polarimeters (SWIRP) project out of NASA’s Goddard Spaceflight Center. The IRCSP scientific objective targets measurements of AOLP and DOLP with 1-μm spectral resolution from 8.5 - 12.5 μm in a single snapshot. The geometry of the field stop determines the field of view (FOV) of the IRCSP. This work relates the spectral resolution, instrument throughput, and polarimetric accuracy of a spectro-polarimeter to the FOV. The accuracy of linear Stokes retrievals for low temperature thermal targets are predicted for varying FOV and measurement noise conditions. This work presents a method to quantify the achievable accuracy in AOLP and DOLP as a function of field stop dimensions and signal-to-noise ratio (SNR). While smaller field stops are shown to improve accuracy as the spectral resolution is increased, low SNR is the dominant source of error for the IRCSP prototype. For the IRCSP, a SNR of at least 80 is required to produce DOLP measurements with < 5% error for targets with DOLP < 0.2.

LWIR↗