MODIS Solar Diffuser Degradation Determination and Its Spectral Dependency

Abstract

This study presents a modeling approach to improve solar diffuser (SD) degradation determination from SD stability monitor (SDSM) measurements. The MODIS instrument uses a SDto calibrate its reflective solar bands (RSBs) on-orbit. Due to the imperfectly designed SDSM sun view screen, the SD reflectance tracked by SDSM has large noise. The SDSM measurements noise is spectrally coherent and can be minimized by normalizing measurements to the least degradeddetector 9 (936 nm). In this study, a SD degradation model is used to determine the SDdegradation's wavelength dependency and the detector 9 degradation is estimated by the model solution.The results show the SD degradations measured at 6 SDSM detectors (554 _ 936 nm) have stable relationships, where the degradation is inversely proportion to 1/wavelength^4. The model estimated SD degradation at SDSM detector 9 wavelength (936 nm) is ~0.9% from 2002 to 2018.Based on the SD degradation model solution, the SD degradation at short/mid wave bands are estimated to improve short/mid wave bands calibration. The model can also be used to improve interpolating SD degradation at SDSM detectors to RSB wavelengths. Compared to linear interpolation, bands 9 and 10 show the largest differences of up to 0.3 and 0.4% respectively. These differences directly impact the calibration coefficients of these bands.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Lee, Shihyan, Meister, Gerhard. 2018-09-07. MODIS Solar Diffuser Degradation Determination and Its Spectral Dependency. https://ntrs.nasa.gov/citations/20190000955

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Stability Assessment of OCO-2 Radiometric Calibration Using Aqua MODIS as a Reference

With three imaging grating spectrometers, the Orbiting Carbon Observatory-2 (OCO-2)measures high spectral resolution spectra (λ/∆λ≈19,000) of reflected solar radiation within the molecular oxygen (O2) A-band at 0.765 μm and two carbon dioxide (CO2) bands at 1.61 and 2.06 μm. OCO-2 uses onboard lamps with a reflective diffuser, solar observations through a transmissive diffuser, lunar measurements, and surface targets for radiometric calibration and validation. Separating calibrator aging from instrument degradation poses a challenge to OCO-2. Here we present a methodology for trending the OCO-2 Build 8R radiometric calibration using OCO-2 nadir observations over eight desert sites and nearly simultaneous observations from Moderate Resolution Imaging Spectroradiometer (MODIS) with sensor viewing zenith angles of 15 ± 0.5°. For the O2 A-band, this methodology is able to quantify a drift of−0.8 ± 0.1% per year and capture a small error in correcting the aging of the solar calibrator. For the other two OCO-2 bands, no measurable changes were seen, indicating less than 0.1% and less than 0.3% per year drift in the radiometric calibration of Band 2 and Band 3, respectively.

radiometric calibration↗

Initial Investigation of the Angular Dependence of the NOAA-20 VIIRS Solar Diffuser BRDF Change Factor

The NOAA-20 (formerly the Joint Polar Satellite System-1) satellite was launched on November 18, 2017. One of the five scientific instruments aboard the NOAA-20 satellite (N20) is the Visible Infrared Imaging Radiometer Suite (VIIRS). The VIIRS scans the earth surface in 22 spectral bands, of which 14 are denoted as the reflective solar bands (RSBs) with design band central wavelengths from 412 to 2250 nm. The VIIRS regularly performs on-orbit radiometric calibration of its RSBs, primarily through observations of an onboard sunlit solar diffuser (SD). The on-orbit change of the SD bidirectional reflectance distribution function (BRDF) value, denoted as the H-factor, is determined by an onboard solar diffuser stability monitor (SDSM). We have shown that the H-factor for the SD on the VIIRS instrument on the Suomi National Polar-orbiting Partnership (SNPP) satellite is both incident and outgoing sunlight direction dependent. This angular dependence profoundly affects the on-orbit radiometric calibration process and results. Here, we give preliminary results for the angular dependence for the N20 VIIRS SD H-factor, and compare the dependence with that for the SNPP VIIRS.

radiometric calibration↗

Absolute Calibration of Optical Satellite Sensors Using Libya 4 Pseudo Invariant Calibration Site

The objective of this paper is to report the improvements in an empirical absolute calibration model developed at South Dakota State University using Libya 4 (+28.55 deg, +23.39 deg) pseudo invariant calibration site (PICS). The approach was based on use of the Terra MODIS as the radiometer to develop an absolute calibration model for the spectral channels covered by this instrument from visible to shortwave infrared. Earth Observing One (EO-1) Hyperion, with a spectral resolution of 10 nm, was used to extend the model to cover visible and near-infrared regions. A simple Bidirectional Reflectance Distribution function (BRDF) model was generated using Terra Moderate Resolution Imaging Spectroradiometer (MODIS) observations over Libya 4 and the resulting model was validated with nadir data acquired from satellite sensors such as Aqua MODIS and Landsat 7 (L7) Enhanced Thematic Mapper (ETM+). The improvements in the absolute calibration model to account for the BRDF due to off-nadir measurements and annual variations in the atmosphere are summarized. BRDF models due to off-nadir viewing angles have been derived using the measurements from EO-1 Hyperion. In addition to L7 ETM+, measurements from other sensors such as Aqua MODIS, UK-2 Disaster Monitoring Constellation (DMC), ENVISAT Medium Resolution Imaging Spectrometer (MERIS) and Operational Land Imager (OLI) onboard Landsat 8 (L8), which was launched in February 2013, were employed to validate the model. These satellite sensors differ in terms of the width of their spectral bandpasses, overpass time, off-nadir-viewing capabilities, spatial resolution and temporal revisit time, etc. The results demonstrate that the proposed empirical calibration model has accuracy of the order of 3% with an uncertainty of about 2% for the sensors used in the study.

radiometric calibration↗