DOE OSTI · 1004951
Temperature Control Diagnostics for Sample Environments
Abstract
In a scientific laboratory setting, standard equipment such as cryocoolers are often used as part of a custom sample environment system designed to regulate temperature over a wide range. The end user may be more concerned with precise sample temperature control than with base temperature. But cryogenic systems tend to be specified mainly in terms of cooling capacity and base temperature. Technical staff at scientific user facilities (and perhaps elsewhere) often wonder how to best specify and evaluate temperature control capabilities. Here we describe test methods and give results obtained at a user facility that operates a large sample environment inventory. Although this inventory includes a wide variety of temperature, pressure, and magnetic field devices, the present work focuses on cryocooler-based systems.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Santodonato, Louis J, Walker, Lakeisha MH, Church, Andrew J, Redmon, Christopher Mckenzie. 2010-01-01. Temperature Control Diagnostics for Sample Environments. https://www.osti.gov/biblio/1004951