DOE OSTI · 1171122
Coded Aperture Imaging for Fluorescent X-rays-Biomedical Applications
Abstract
Employing a coded aperture pattern in front of a charge couple device pixilated detector (CCD) allows for imaging of fluorescent x-rays (6-25KeV) being emitted from samples irradiated with x-rays. Coded apertures encode the angular direction of x-rays and allow for a large Numerical Aperture x- ray imaging system. The algorithm to develop the self-supported coded aperture pattern of the Non Two Holes Touching (NTHT) pattern was developed. The algorithms to reconstruct the x-ray image from the encoded pattern recorded were developed by means of modeling and confirmed by experiments. Samples were irradiated by monochromatic synchrotron x-ray radiation, and fluorescent x-rays from several different test metal samples were imaged through the newly developed coded aperture imaging system. By choice of the exciting energy the different metals were speciated.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Haboub, Abdel, MacDowell, Alastair, Marchesini, Stefano, Parkinson, Dilworth. 2013-06-01. Coded Aperture Imaging for Fluorescent X-rays-Biomedical Applications. https://www.osti.gov/biblio/1171122
Cite the original work for its findings. Save a collection to share your selection of sources.