DOE OSTI · 1184089
Neutron Sources for Standard-Based Testing
Abstract
The DHS TC Standards and the consensus ANSI Standards use 252 Cf as the neutron source for performance testing because its energy spectrum is similar to the 235 U and 239 Pu fission sources used in nuclear weapons. An emission rate of 20,000 ± 20% neutrons per second is used for testing of the radiological requirements both in the ANSI standards and the TCS. Determination of the accurate neutron emission rate of the test source is important for maintaining consistency and agreement between testing results obtained at different testing facilities. Several characteristics in the manufacture and the decay of the source need to be understood and accounted for in order to make an accurate measurement of the performance of the neutron detection instrument. Additionally, neutron response characteristics of the particular instrument need to be known and taken into account as well as neutron scattering in the testing environment.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Radev, Radoslav, McLean, Thomas. 2014-11-10. Neutron Sources for Standard-Based Testing. https://doi.org/10.2172/1184089
Cite the original work for its findings. Save a collection to share your selection of sources.