Active Well Counting Using New PSD Plastic Detectors

Abstract

This report presents results and analysis from a series of proof-of-concept measurements to assess the suitability of segmented detectors constructed from Eljen EJ-299-34 PSD-plastic scintillator with pulse-shape discrimination capability for the purposes of quantifying uranium via active neutron coincidence counting. Present quantification of bulk uranium materials for international safeguards and domestic materials control and accounting relies on active neutron coincidence counting systems, such as the Active Well Coincidence Counter (AWCC) and the Uranium Neutron Coincidence Collar (UNCL), that use moderated He-3 proportional counters along with necessarily low-intensity 241Am(Li) neutron sources. Scintillation-based fast-neutron detectors are a potentially superior technology to the existing AWCC and UNCL designs due to their spectroscopic capability and their inherently short neutron coincidence times that largely eliminate random coincidences and enable interrogation by stronger sources. One of the past impediments to the investigation and adoption of scintillation counters for the purpose of quantifying bulk uranium was the commercial availability of scintillators having the necessary neutron-gamma pulse-shape discrimination properties only as flammable liquids. Recently, Eljen EJ-299-34 PSD-plastic scintillator became commercially available. The present work is the first assessment of an array of PSD-plastic detectors for the purposes of quantifying bulk uranium. The detector panel used in the present work was originally built as the focal plane for a fast-neutron imager, but it was repurposed for the present investigation by construction of a stand to support the inner well of an AWCC immediately in front of the detector panel. The detector panel and data acquisition of this system are particularly well suited for performing active-well fast-neutron counting of LEU and HEU samples because the active detector volume is solid, the 241Am(Li) interrogating neutrons are largely below the detector threshold, and the segmented construction of the detector modules allow for separation of true neutron-neutron coincidences from inter-detector scattering using the kinematics of neutron scattering. The results from a series of measurements of a suite of uranium standards are presented, and compared to measurements of the same standards and source configurations using the AWCC. Using these results, the performance of the segmented detectors reconfigured as a well counter is predicted and outperforms the AWCC.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Hausladen, Paul, Newby, Jason, McElroy, Robert Dennis. 2015-11-01. Active Well Counting Using New PSD Plastic Detectors. https://doi.org/10.2172/1240561. https://searchnasa.com/record/osti%3A1240561/active-well-counting-using-new-psd-plastic-detectors

Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search NASA by Indiciae: BibTeX · RIS.

KEEP EXPLORING

Related reports

Pulsed-Neutron Experiments at the Inherently Safe Subcritical Assembly

The pulsed neutron technique is a powerful, dynamic method to assay the reactivity of a multiplying system. This work presents the novel application of the pulsed neutron technique to the Inherently Safe Subcritical Assembly, an experimental configuration accepted by the International Criticality Safety Benchmark Evaluation Project Handbook. The experiments were replicated with COG11.3, a continuous-energy Monte Carlo code. The pulsed neutron data were analyzed using the Sjöstrand and Gozani area-ratio methods and by extracting the prompt neutron decay constant. Subsequent static k-eigenvalue and 𝛼-eigenvalue simulations were also performed for the same configurations. Neutron detector dead-time effects from the experiments were corrected using the Backwards Extrapolation Method and shown to have a negligible impact on the estimated reactivities. The results highlight that capturing time-dependent effects like delayed neutron precursor buildup are essential to accurately reproduce experimental results. They also highlight the importance of shielding the detectors from generator source neutrons in deeply subcritical configurations.

21 SPECIFIC NUCLEAR REACTORS AND ASSOCIATED PLANTS↗

Machine Learning–Based Condition Monitoring of a Circulating Water System of a Canadian Nuclear Plant

With the need to maintain long-term reliable energy using nuclear power plants, there is an underlying demand to ensure that the maintenance of plant components and systems is also done in an efficient and cost-effective manner. One way to achieve this is by moving from time-based maintenance to condition-based maintenance. The research presented in this paper focuses on applying statistical and machine-learning-based methods to capture anomalies within data for fault detection to further develop into condition monitoring. This paper focuses on system data for a circulating water system (CWS) of a pressurized heavy-water reactor for detecting anomalies. The different methodologies used for detecting and capturing anomalies in the CWS data are matrix profile, density-based spatial clustering of applications with noise (DBSCAN), and support vector machines (SVMs). Matrix profile and DBSCAN are used to distinguish between normal data and anomalous data. This paper presents a hybrid method using DBSCAN and SVM when a portion of the data is used for DBSCAN to generate clusters. This portion of data is then used to train the SVM along with the clusters generated by DBSCAN as output. SVM is then tested on unseen data as a predictive tool, which can work in real time to categorize data points as either normal or anomalous. This paper presents results that show the high accuracies of DBSCAN and SVM in capturing anomalies within the data for a CWS for fault detection. Thus, the maintenance plan would be focused on component condition rather than a time-based schedule by switching to an automated system to identify and predict faults within a CWS.

21 SPECIFIC NUCLEAR REACTORS AND ASSOCIATED PLANTS↗

SCALE HTR-PROTEUS Benchmark Model

This dataset contains input and result files of computational simulations of HTR-PROTEUS benchmark with the latest version of SCALE code system. The simulations cover criticality control rod worth calculations as well as sensitivity analysis and uncertainty quantification. Users wanting to reproduce results from this dataset are required to obtain a license to the SCALE code system for which details on the distribution can be found here: https://www.ornl.gov/scale/releases

21 SPECIFIC NUCLEAR REACTORS AND ASSOCIATED PLANTS↗