Digital pulse processing for NaI(Tl) detectors

Abstract

We apply two different post-processing techniques to digital pulses induced by photons in a NaI(Tl) detector and compare the obtained energy resolution to the standard analog approach. Our digital acquisition approach is performed using a single-stage acquisition with a fast digitizer. Both the post-processing techniques we propose rely on signal integration. In the first, the pulse integral is calculated by directly numerically integrating the pulse digital samples, while in the second the pulse integral is estimated by a model-based fitting of the pulse. Our study used a 7.62 cm×7.62 cm cylindrical NaI(Tl) detector that gave a 7.60% energy resolution (at 662 keV), using the standard analog acquisition approach, based on a pulse shaping amplifier. The new direct numerical integration yielded a 6.52% energy resolution. The fitting approach yielded a 6.55% energy resolution, and, although computationally heavier than numerical integration, is preferable when only the early samples of the pulse are available. We also evaluated the timing performance of a fast–slow detection system, encompassing an EJ-309 and a NaI(Tl) scintillator. Here, we use two techniques to determine the pulse start time: constant fraction discrimination (CFD) and adaptive noise threshold timing (ANT), for both the analog and digital acquisition approach. With the analog acquisition approach, we found a system time resolution of 5.8 ns and 7.3 ns, using the constant fraction discrimination and adaptive noise threshold timing, respectively. With the digital acquisition approach, a time resolution of 1.2 ns was achieved using the ANT method and 3.3 ns using CFD at 50% of the maximum, to select the pulse start time. Finally, the proposed direct digital readout and post-processing techniques can improve the application of NaI(Tl) detectors, traditionally considered ‘slow’, for fast counting and correlation measurements, while maintaining a good measurement of the energy resolution.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Di Fulvio, A., Shin, T. H., Hamel, M. C., Pozzi, S. A.. 2015-10-02. Digital pulse processing for NaI(Tl) detectors. https://doi.org/10.1016/j.nima.2015.09.080. https://searchnasa.com/record/osti%3A1365802/digital-pulse-processing-for-nai-tl-detectors

Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search NASA by Indiciae: BibTeX · RIS.

KEEP EXPLORING

Related reports

Performance evaluation of neutron noise analysis in detecting special nuclear materials

Reliable and rapid inspection techniques play a vital role in preventing illicit trafficking of special nuclear materials. Active interrogation systems using neutrons produced by portable, high-flux deuterium-deuterium or deuterium-tritium neutron generators are being actively developed as a secondary scanning tool for this purpose. In this study, a neutron noise analysis-based approach for detecting unshielded and shielded special nuclear materials by using a pulsed deuterium-tritium neutron generator was evaluated. Here, this approach analyzes the fluctuation of neutron counts. Its performance was quantified with regard to time-to-detection to achieve a minimum probability of detection of 99% and a probability of false alarm of less than 1% considering various amounts of special nuclear materials and different shielding configurations. It was demonstrated that this approach could detect 17 uranium slugs in 5 s given a neutron generator yield of 8.1 × 10 7 n/s. These slugs could be detected within a reasonable time frame (200 s) when they were shielded by 10.16 cm of high-density polyethylene. The results obtained using the neutron noise analysis approach were compared with those obtained using the commonly used differential die-away analysis technique, a sensitive technique for detecting the presence of fissile materials by utilizing the prompt fission neutrons produced when the source neutrons from a neutron generator are completely diminished. For example, the time to detect 2 unshielded uranium slugs was 2.1 s when using the differential die-away analysis technique; it increased to 93 s for the neutron noise analysis approach. Although the noise analysis-based approach exhibits an overall performance which is not as good as that of differential die-away, neutron noise provides an alternative method for effective detection of special nuclear materials.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

A proximity-focusing RICH detector for the ePIC Experiment at the EIC

The Electron-Proton/Ion Collider Experiment (ePIC) will be a large, multi-purpose detector to be installed at the Electron-Ion Collider (EIC) being built at Brookhaven National Laboratory. As robust particle identification (PID) capabilities are essential for fully realizing the EIC science program, ePIC contains several PID subsystems spanning different angular ranges. PID capability in the electron-going endcap is provided by a proximity-focusing Ring Imaging Cherenkov detector (pfRICH) designed to deliver at least 3σ separation between pions and kaons for momenta up to 7 GeV/ c . It will also aid with electron-hadron discrimination at low momentum and assist in the determination of the collision time (t 0 ). This contribution will summarize the design of the pfRICH as well as ongoing fabrication and component testing efforts. geant-based performance simulations will also be discussed.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Design of a high voltage delivery system for noble liquid time projection chambers

Noble liquid time projection chambers (TPCs) are a leading technology in the detection of ionizing radiation, particularly in applications such as accelerator neutrino physics, dark matter detection, and neutrinoless double beta decay. This work addresses the design considerations for implementing stable high voltage (HV) systems within large noble liquid TPCs, with a focus on the nEXO experiment. Utilizing insights from prior HV research and experimental investigations, we outline factors influencing HV stability and discuss design choices to improve stability and prevent electrical discharges. A novel HV delivery system concept is presented, tailored for the nEXO TPC, which incorporates these design considerations while also meeting the stringent radiopurity requirements of the nEXO neutrinoless double beta decay search. These design considerations and their specific implementation towards a HV delivery system offer guidance to future experiments applying high voltage in noble liquid environments.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗