MHD-240 Neutron Detection Efficiency and Sensitivity Calibration
Abstract
The neutron detection efficiency of the MHD-240 scintillation detector was measured at Ohio University Edwards Accelerator Laboratory. Pulsed broad-spectrum sources from the 27 Al(d,n) and 9 Be(d,n) reactions were used to produce the neutrons. The neutron time-of-flight technique was used to make an energy-differential efficiency measurement. The measurements were made in a low rate pulse mode. The pulse height distribution was also recorded, and the data were used to derive a neutron sensitivity that would be applicable in a high rate environment such as a critical assembly where the detector would be expected to operate in a current mode.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Snyder, Lucas A., Bower, Dan E., Fittinghoff, David N., May, Mark J.. 2020-03-12. MHD-240 Neutron Detection Efficiency and Sensitivity Calibration. https://doi.org/10.2172/1605527