Abstract
In laboratory plasmas, arrays of probes have typically been used to measure the evolution of the magnetic field topology. Here, we present initial image-based measurements of the magnetic topology in a low-temperature plasma using a purely optical diagnostic. Laser induced fluorescence measurements of neutral velocity distribution functions are made using a fast camera, imaging the Zeeman splitting of σ-peaks in neutral argon. The separation of σ-peaks provides spatially resolved magnetic field magnitude measurements with a detection threshold on the order of 10 G
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Gilbert, T. J., Stevenson, K. J., Paul, M. C., Steinberger, T. E., Scime, E. E.. 2021-05-12. Magnetic field imaging in a laboratory plasma. https://doi.org/10.1063/5.0052957