DOE OSTI · 1969760
A tabletop X-ray tomography instrument for nanometer-scale imaging: reconstructions
Abstract
Abstract We show three-dimensional reconstructions of a region of an integrated circuit from a 130 nm copper process. The reconstructions employ x-ray computed tomography, measured with a new and innovative high-magnification x-ray microscope. The instrument uses a focused electron beam to generate x-rays in a 100 nm spot and energy-resolving x-ray detectors that minimize backgrounds and hold promise for the identification of materials within the sample. The x-ray generation target, a layer of platinum, is fabricated on the circuit wafer itself. A region of interest is imaged from a limited range of angles and without physically removing the region from the larger circuit. The reconstruction is consistent with the circuit’s design file.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Levine, Zachary H. (ORCID:000000029928276X), Alpert, Bradley K. (ORCID:0000000197659642), Dagel, Amber L. (ORCID:0000000258821640), Fowler, Joseph W., Jimenez, Edward S., Nakamura, Nathan, Swetz, Daniel S., Szypryt, Paul, Thompson, Kyle R., Ullom, Joel N.. 2023-04-14. A tabletop X-ray tomography instrument for nanometer-scale imaging: reconstructions. https://doi.org/10.1038/s41378-023-00510-6