Abstract

Purpose: For the last few years, development and optimization of three-dimensional (3D) x-ray breast imaging systems, such as digital breast tomosynthesis (DBT) and computed tomography, have drawn much attention from the medical imaging community, either academia or industry. However, there is still much room for understanding how to best optimize and evaluate the devices over a large space of many different system parameters and geometries. Current evaluation methods, which work well for 2D systems, do not incorporate the depth information from the 3D imaging systems. Therefore, it is critical to develop a statistically sound evaluation method to investigate the usefulness of inclusion of depth and background-variability information into the assessment and optimization of the 3D systems. Methods: In this paper, we present a mathematical framework for a statistical assessment of planar and 3D x-ray breast imaging systems. Our method is based on statistical decision theory, in particular, making use of the ideal linear observer called the Hotelling observer. We also present a physical phantom that consists of spheres of different sizes and materials for producing an ensemble of randomly varying backgrounds to be imaged for a given patient class. Lastly, we demonstrate our evaluation method in comparing laboratory mammography and three-angle DBT systems for signal detection tasks using the phantom's projection data. We compare the variable phantom case to that of a phantom of the same dimensions filled with water, which we call the uniform phantom, based on the performance of the Hotelling observer as a function of signal size and intensity. Results: Detectability trends calculated using the variable and uniform phantom methods are different from each other for both mammography and DBT systems. Conclusions: Our results indicate that measuring the system's detection performance with consideration of background variability may lead to differences in system performance estimates and comparisons. For the assessment of 3D systems, to accurately determine trade offs between image quality and radiation dose, it is critical to incorporate randomness arising from the imaging chain including background variability into system performance calculations.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Park, Subok, Jennings, Robert, Haimo, Liu, Badano, Aldo, Myers, Kyle. 2010-12-15. A statistical, task-based evaluation method for three-dimensional x-ray breast imaging systems using variable-background phantoms. https://doi.org/10.1118/1.3488910

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Cross sections of 147–149 Sm( 6 Li,x) reactions for the production of 149 Tb for targeted alpha therapy

Terbium-149g (t 1/2 = 4.12 h) is of particular interest for targeted alpha therapy cancer treatment due to its ability to decay via both alpha and positron emission, making it a potential theranostic nuclide. Due to many challenges facing its production, there are limited facilities worldwide that have demonstrated the ability to produce this nuclide in quantities sufficient for medical research. Since the Cyclotron Institute at Texas A&M University is a specialized accelerator facility capable of accelerating a wide variety of ions, we are investigating production pathway options. One of the major challenges facing its production is the known co-production of the excited isomeric state, 149m Tb (t 1/2 = 4.1 min). However, this state does not decay to the ground state of 149g Tb, negating any potential contribution to its yield. Due to its short-half life, the cross section for the population of this state has never been measured. After calculating several potential reaction yields using predictive models, the reactions of 147–149 Sm( 6 Li,xn) 149 Tb were identified as candidates. Lithium-6 beams of varied energies between 45-65 MeV were impinged on enriched 147 Sm, 148 Sm, and 149 Sm targets at the Cyclotron Institute at Texas A&M University, and the reaction products were measured immediately following irradiation using high-purity germanium detectors, enabling detection of both 149m Tb and 149g Tb. Cross sections for all nuclides produced in sufficient activity in these reactions were also measured and reported here. We conclude that the population of 149m Tb is much preferred over population of the ground state for these 6 Li-induced reactions, and it is necessary to explore other options for 149g Tb production.

62 RADIOLOGY AND NUCLEAR MEDICINE↗

Source function from two-particle correlation function through entropy-regularized Richardson-Lucy deblurring

Source functions are obtained from p – p and d – α correlation functions by applying the Richardson-Lucy (RL) deblurring to the Koonin-Pratt (KP) equation. To prevent fitting of noise in the correlation function, total-variation (TV) regularization is employed that has been effective in ordinary image restoration. TV alone cannot ensure normalization of the source functions. To ensure the latter, we propose a maximum-entropy regularized RL algorithm (MEM-RL). We outline the MEM-RL formalism and optimization strategy for the KP equation, demonstrating its effectiveness on both simulated and experimental data, including the p – p and d – α correlation functions.

62 RADIOLOGY AND NUCLEAR MEDICINE↗