A software-based x-ray scatter correction method for breast tomosynthesis

Abstract

Purpose: To develop a software-based scatter correction method for digital breast tomosynthesis (DBT) imaging and investigate its impact on the image quality of tomosynthesis reconstructions of both phantoms and patients. Methods: A Monte Carlo (MC) simulation of x-ray scatter, with geometry matching that of the cranio-caudal (CC) view of a DBT clinical prototype, was developed using the Geant4 toolkit and used to generate maps of the scatter-to-primary ratio (SPR) of a number of homogeneous standard-shaped breasts of varying sizes. Dimension-matched SPR maps were then deformed and registered to DBT acquisition projections, allowing for the estimation of the primary x-ray signal acquired by the imaging system. Noise filtering of the estimated projections was then performed to reduce the impact of the quantum noise of the x-ray scatter. Three dimensional (3D) reconstruction was then performed using the maximum likelihood-expectation maximization (MLEM) method. This process was tested on acquisitions of a heterogeneous 50/50 adipose/glandular tomosynthesis phantom with embedded masses, fibers, and microcalcifications and on acquisitions of patients. The image quality of the reconstructions of the scatter-corrected and uncorrected projections was analyzed by studying the signal-difference-to-noise ratio (SDNR), the integral of the signal in each mass lesion (integrated mass signal, IMS), and the modulation transfer function (MTF). Results: The reconstructions of the scatter-corrected projections demonstrated superior image quality. The SDNR of masses embedded in a 5 cm thick tomosynthesis phantom improved 60%-66%, while the SDNR of the smallest mass in an 8 cm thick phantom improved by 59% (p < 0.01). The IMS of the masses in the 5 cm thick phantom also improved by 15%-29%, while the IMS of the masses in the 8 cm thick phantom improved by 26%-62% (p < 0.01). Some embedded microcalcifications in the tomosynthesis phantoms were visible only in the scatter-corrected reconstructions. The visibility of the findings in two patient images was also improved by the application of the scatter correction algorithm. The MTF of the images did not change after application of the scatter correction algorithm, indicating that spatial resolution was not adversely affected. Conclusions: Our software-based scatter correction algorithm exhibits great potential in improving the image quality of DBT acquisitions of both phantoms and patients. The proposed algorithm does not require a time-consuming MC simulation for each specific case to be corrected, making it applicable in the clinical realm.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Jia Feng, Steve Si, Sechopoulos, Ioannis. 2011-12-15. A software-based x-ray scatter correction method for breast tomosynthesis. https://doi.org/10.1118/1.3659703

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Cross sections of 147–149 Sm( 6 Li,x) reactions for the production of 149 Tb for targeted alpha therapy

Terbium-149g (t 1/2 = 4.12 h) is of particular interest for targeted alpha therapy cancer treatment due to its ability to decay via both alpha and positron emission, making it a potential theranostic nuclide. Due to many challenges facing its production, there are limited facilities worldwide that have demonstrated the ability to produce this nuclide in quantities sufficient for medical research. Since the Cyclotron Institute at Texas A&M University is a specialized accelerator facility capable of accelerating a wide variety of ions, we are investigating production pathway options. One of the major challenges facing its production is the known co-production of the excited isomeric state, 149m Tb (t 1/2 = 4.1 min). However, this state does not decay to the ground state of 149g Tb, negating any potential contribution to its yield. Due to its short-half life, the cross section for the population of this state has never been measured. After calculating several potential reaction yields using predictive models, the reactions of 147–149 Sm( 6 Li,xn) 149 Tb were identified as candidates. Lithium-6 beams of varied energies between 45-65 MeV were impinged on enriched 147 Sm, 148 Sm, and 149 Sm targets at the Cyclotron Institute at Texas A&M University, and the reaction products were measured immediately following irradiation using high-purity germanium detectors, enabling detection of both 149m Tb and 149g Tb. Cross sections for all nuclides produced in sufficient activity in these reactions were also measured and reported here. We conclude that the population of 149m Tb is much preferred over population of the ground state for these 6 Li-induced reactions, and it is necessary to explore other options for 149g Tb production.

62 RADIOLOGY AND NUCLEAR MEDICINE↗

Source function from two-particle correlation function through entropy-regularized Richardson-Lucy deblurring

Source functions are obtained from p – p and d – α correlation functions by applying the Richardson-Lucy (RL) deblurring to the Koonin-Pratt (KP) equation. To prevent fitting of noise in the correlation function, total-variation (TV) regularization is employed that has been effective in ordinary image restoration. TV alone cannot ensure normalization of the source functions. To ensure the latter, we propose a maximum-entropy regularized RL algorithm (MEM-RL). We outline the MEM-RL formalism and optimization strategy for the KP equation, demonstrating its effectiveness on both simulated and experimental data, including the p – p and d – α correlation functions.

62 RADIOLOGY AND NUCLEAR MEDICINE↗