Search NASA⌕ Search

DOE OSTI · 22098794

Computerized scheme for evaluating mammographic phantom images

Abstract

Purpose: The authors developed a computer algorithm to automatically evaluate images of the American College of Radiology (ACR) mammography accreditation phantom. Methods: The developed algorithm consist of the edge detection of wax insert, nonuniformity correction of background, and correction for magnification and also calculate the cross-correlation coefficient by image matching technique. The algorithm additionally evaluates target shape for fibers, target contrast for speck groups, and target circularity for masses. To obtain an ideal template image without noise and spatial resolution loss, the wax insert containing the embedded test pattern was extracted from the phantom and radiographed. Two template images and ten test phantom images were prepared for this study. The results of evaluation using the algorithm outputs were compared with the averaged results of observer studies by six skilled observers. Results: In comparing the results from the algorithm outputs with the results of observers, the authors found that the computer outputs were well correlated with the evaluations by observers, and they indicate the quality of the phantom image. The correlation coefficients between results of observer studies and two outputs of computer algorithm, i.e., the cross-correlation coefficient by template matching and indices of target shape for fibers, were 0.89 (95% confidence interval, 0.82-0.93; hereinafter the same) and 0.85 (0.76-0.91). The correlation coefficients between observer's results and two outputs: the cross-correlation coefficient and indices of target contrast for speck groups, were 0.83 (0.79-0.86) and 0.85 (0.81-0.88) and between observer's results and two outputs: the cross-correlation coefficient and indices of target circularity for masses, were 0.90 (0.84-0.94) and 0.87 (0.77-0.92). Conclusions: Image evaluation using the ACR phantom is indispensable in quality control of a mammography system. The proposed algorithm is useful for quality control and image evaluation of mammography units.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Asahara, Masaki, Kodera, Yoshie. 2012-03-15. Computerized scheme for evaluating mammographic phantom images. https://doi.org/10.1118/1.3687159. https://searchnasa.com/record/osti%3A22098794/computerized-scheme-for-evaluating-mammographic-phantom-images

Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search NASA by Indiciae: BibTeX · RIS.

KEEP EXPLORING

Related reports

Cross sections of 147–149 Sm( 6 Li,x) reactions for the production of 149 Tb for targeted alpha therapy

Terbium-149g (t 1/2 = 4.12 h) is of particular interest for targeted alpha therapy cancer treatment due to its ability to decay via both alpha and positron emission, making it a potential theranostic nuclide. Due to many challenges facing its production, there are limited facilities worldwide that have demonstrated the ability to produce this nuclide in quantities sufficient for medical research. Since the Cyclotron Institute at Texas A&M University is a specialized accelerator facility capable of accelerating a wide variety of ions, we are investigating production pathway options. One of the major challenges facing its production is the known co-production of the excited isomeric state, 149m Tb (t 1/2 = 4.1 min). However, this state does not decay to the ground state of 149g Tb, negating any potential contribution to its yield. Due to its short-half life, the cross section for the population of this state has never been measured. After calculating several potential reaction yields using predictive models, the reactions of 147–149 Sm( 6 Li,xn) 149 Tb were identified as candidates. Lithium-6 beams of varied energies between 45-65 MeV were impinged on enriched 147 Sm, 148 Sm, and 149 Sm targets at the Cyclotron Institute at Texas A&M University, and the reaction products were measured immediately following irradiation using high-purity germanium detectors, enabling detection of both 149m Tb and 149g Tb. Cross sections for all nuclides produced in sufficient activity in these reactions were also measured and reported here. We conclude that the population of 149m Tb is much preferred over population of the ground state for these 6 Li-induced reactions, and it is necessary to explore other options for 149g Tb production.

62 RADIOLOGY AND NUCLEAR MEDICINE↗

Source function from two-particle correlation function through entropy-regularized Richardson-Lucy deblurring

Source functions are obtained from p – p and d – α correlation functions by applying the Richardson-Lucy (RL) deblurring to the Koonin-Pratt (KP) equation. To prevent fitting of noise in the correlation function, total-variation (TV) regularization is employed that has been effective in ordinary image restoration. TV alone cannot ensure normalization of the source functions. To ensure the latter, we propose a maximum-entropy regularized RL algorithm (MEM-RL). We outline the MEM-RL formalism and optimization strategy for the KP equation, demonstrating its effectiveness on both simulated and experimental data, including the p – p and d – α correlation functions.

62 RADIOLOGY AND NUCLEAR MEDICINE↗