DOE OSTI · 22202367
A poly(dimethylsiloxane)-based device enabling time-lapse imaging with high spatial resolution
Abstract
We have developed a regulator-free device that enables long-term incubation of mammalian cells for epi-fluorescence imaging, based on a concept that the size of sample to be gassed and heated is reduced to observation scale. A poly(dimethylsiloxane) block stamped on a coverslip works as a long-lasting supplier of CO{sub 2}-rich gas to adjust bicarbonate-containing medium in a tiny chamber at physiological pH, and an oil-immersion objective warms cells across the coverslip. A time-lapse imaging experiment using HeLa cells stably expressing fluorescent cell-cycle indicators showed that the cells in the chamber proliferated with normal cell-cycle period over 2 days.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Hirano, Masahiko, Hoshida, Tetsushi, Sakaue-Sawano, Asako, Miyawaki, Atsushi. 2010-02-12. A poly(dimethylsiloxane)-based device enabling time-lapse imaging with high spatial resolution. https://doi.org/10.1016/j.bbrc.2009.12.181. https://searchnasa.com/record/osti%3A22202367/a-poly-dimethylsiloxane-based-device-enabling-time-lapse-imaging-with-high-spatial-resolution
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search NASA by Indiciae: BibTeX · RIS.