Search NASA⌕ Search

DOE OSTI · 22493983

SU-E-I-24: Method for CT Automatic Exposure Control Verification

Abstract

Purpose: Design of a phantom and a simple method for the automatic exposure control (AEC) verification in CT. This verification is included in the computed tomography (CT) Spanish Quality Assurance Protocol. Methods: The phantom design is made from the head and the body phantom used for the CTDI measurement and PMMA plates (35×35 cm2) of 10 cm thickness. Thereby, three different thicknesses along the longitudinal axis are obtained which permit to evaluate the longitudinal AEC performance. Otherwise, the existent asymmetry in the PMMA layers helps to assess angular and 3D AEC operation.Recent acquisition in our hospital (August 2014) of Nomex electrometer (PTW), together with the 10 cm pencil ionization chamber, led to register dose rate as a function of time. Measurements with this chamber fixed at 0° and 90° on the gantry where made on five multidetector-CTs from principal manufacturers. Results: Individual analysis of measurements shows dose rate variation as a function of phantom thickness. The comparative analysis shows that dose rate is kept constant in the head and neck phantom while the PMMA phantom exhibits an abrupt variation between both results, being greater results at 90° as the thickness of the phantom is 3.5 times larger than in the perpendicular direction. Conclusion: Proposed method is simple, quick and reproducible. Results obtained let a qualitative evaluation of the AEC and they are consistent with the expected behavior. A line of future development is to quantitatively study the intensity modulation and parameters of image quality, and a possible comparative study between different manufacturers.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Gracia, M, Olasolo, J, Martin, M, Bragado, L, Gallardo, N, Miquelez, S, Maneru, F, Lozares, S, Pellejero, S, Rubio, A. 2015-06-15. SU-E-I-24: Method for CT Automatic Exposure Control Verification. https://doi.org/10.1118/1.4924021

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Novel Method for Domestic Stable Isotope Production (CRADA #667) Final Report

This report presents the progress achieved on our chlorine isotope separation efforts in the CRADA #667 agreement. We used a microchannel distillation (MCD) packing structure to enrich HCl isotopes (H35Cl, H37Cl) and an isotachophoresis (ITP) separation to enrich ionic chloride species (35Cl- and 37Cl-). A 2 m tall MCD column was constructed and achieved 264 separation stages over several days, which is well above the goal of 50 stages outlined in the CRADA. The resulting height equivalent to a theoretical plate (HETP) was 0.76 cm. ITP tests were run ~ 24 hours and achieved a single-stage separation factor of 1.559.

07 ISOTOPE AND RADIATION SOURCES↗