TU-FG-201-09: Predicting Accelerator Dysfunction

Abstract

Purpose: To develop an integrated statistical process control (SPC) framework using digital performance and component data accumulated within the accelerator system that can detect dysfunction prior to unscheduled downtime. Methods: Seven digital accelerators were monitored for twelve to 18 months. The accelerators were operated in a ‘run to failure mode’ with the individual institutions determining when service would be initiated. Institutions were required to submit detailed service reports. Trajectory and text log files resulting from a robust daily VMAT QA delivery were decoded and evaluated using Individual and Moving Range (I/MR) control charts. The SPC evaluation was presented in a customized dashboard interface that allows the user to review 525 monitored parameters (480 MLC parameters). Chart limits were calculated using a hybrid technique that includes the standard SPC 3σ limits and an empirical factor based on the parameter/system specification. The individual (I) grand mean values and control limit ranges of the I/MR charts of all accelerators were compared using statistical (ranked analysis of variance (ANOVA)) and graphical analyses to determine consistency of operating parameters. Results: When an alarm or warning was directly connected to field service, process control charts predicted dysfunction consistently on beam generation related parameters (BGP)– RF Driver Voltage, Gun Grid Voltage, and Forward Power (W); beam uniformity parameters – angle and position steering coil currents; and Gantry position accuracy parameter: cross correlation max-value. Control charts for individual MLC – cross correlation max-value/position detected 50% to 60% of MLCs serviced prior to dysfunction or failure. In general, non-random changes were detected 5 to 80 days prior to a service intervention. The ANOVA comparison of BGP determined that each accelerator parameter operated at a distinct value. Conclusion: The SPC framework shows promise. Long term monitoring coordinated with service will be required to definitively determine the effectiveness of the model. Varian Medical System, Inc. provided funding in support of the research presented.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Able, C, Nguyen, C, Baydush, A, Munley, M, Gersh, J, Ndlovu, A, Rebo, I, Booth, J, Perez, M, Sintay, B. 2016-06-15. TU-FG-201-09: Predicting Accelerator Dysfunction. https://doi.org/10.1118/1.4957532

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

Performance evaluation of neutron noise analysis in detecting special nuclear materials

Reliable and rapid inspection techniques play a vital role in preventing illicit trafficking of special nuclear materials. Active interrogation systems using neutrons produced by portable, high-flux deuterium-deuterium or deuterium-tritium neutron generators are being actively developed as a secondary scanning tool for this purpose. In this study, a neutron noise analysis-based approach for detecting unshielded and shielded special nuclear materials by using a pulsed deuterium-tritium neutron generator was evaluated. Here, this approach analyzes the fluctuation of neutron counts. Its performance was quantified with regard to time-to-detection to achieve a minimum probability of detection of 99% and a probability of false alarm of less than 1% considering various amounts of special nuclear materials and different shielding configurations. It was demonstrated that this approach could detect 17 uranium slugs in 5 s given a neutron generator yield of 8.1 × 10 7 n/s. These slugs could be detected within a reasonable time frame (200 s) when they were shielded by 10.16 cm of high-density polyethylene. The results obtained using the neutron noise analysis approach were compared with those obtained using the commonly used differential die-away analysis technique, a sensitive technique for detecting the presence of fissile materials by utilizing the prompt fission neutrons produced when the source neutrons from a neutron generator are completely diminished. For example, the time to detect 2 unshielded uranium slugs was 2.1 s when using the differential die-away analysis technique; it increased to 93 s for the neutron noise analysis approach. Although the noise analysis-based approach exhibits an overall performance which is not as good as that of differential die-away, neutron noise provides an alternative method for effective detection of special nuclear materials.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

A proximity-focusing RICH detector for the ePIC Experiment at the EIC

The Electron-Proton/Ion Collider Experiment (ePIC) will be a large, multi-purpose detector to be installed at the Electron-Ion Collider (EIC) being built at Brookhaven National Laboratory. As robust particle identification (PID) capabilities are essential for fully realizing the EIC science program, ePIC contains several PID subsystems spanning different angular ranges. PID capability in the electron-going endcap is provided by a proximity-focusing Ring Imaging Cherenkov detector (pfRICH) designed to deliver at least 3σ separation between pions and kaons for momenta up to 7 GeV/ c . It will also aid with electron-hadron discrimination at low momentum and assist in the determination of the collision time (t 0 ). This contribution will summarize the design of the pfRICH as well as ongoing fabrication and component testing efforts. geant-based performance simulations will also be discussed.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Design of a high voltage delivery system for noble liquid time projection chambers

Noble liquid time projection chambers (TPCs) are a leading technology in the detection of ionizing radiation, particularly in applications such as accelerator neutrino physics, dark matter detection, and neutrinoless double beta decay. This work addresses the design considerations for implementing stable high voltage (HV) systems within large noble liquid TPCs, with a focus on the nEXO experiment. Utilizing insights from prior HV research and experimental investigations, we outline factors influencing HV stability and discuss design choices to improve stability and prevent electrical discharges. A novel HV delivery system concept is presented, tailored for the nEXO TPC, which incorporates these design considerations while also meeting the stringent radiopurity requirements of the nEXO neutrinoless double beta decay search. These design considerations and their specific implementation towards a HV delivery system offer guidance to future experiments applying high voltage in noble liquid environments.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗