Spatial Resolution of a microCT Scanner
Abstract
We have measured the spatial resolution of the 450 kV test-bed microCT scanner. We have also modeled the resolution analytically. The model and measurement show correspondence within 5%. Our measurement method was to use a thin tungsten wire to act as an impulse function. Therefore, the image is the impulse response, also called the point spread function (PSF) of the system. The Fourier transform (FT) of the impulse response is the modulation transfer function (MTF) of the system. We report the MTF here. By moving the wire at different positions in the scan field of view (FOV), we can measure MTF at different positions in the FOV. Since our imaging system is radially symmetric, we report MTF in two directions, radial and tangential as shown in Figure 1. The radial direction is along the line from isocenter to the wire, and the tangential is perpendicular to it.