SIMS Data Correction Procedure for Quasi‐Simultaneous Arrival ( QSA ) Under‐counting and Ramifications of Misapplication
Also available from
Abstract
Isotope geochemistry requires isotope ratios measured using secondary ion mass spectrometry (SIMS) to be made with optimal precision and accuracy. Under some analytical conditions when using electron multiplier detectors, secondary ions may be under‐counted because of quasi‐simultaneous arrival (QSA) at the first dynode. The relative magnitude of the associated QSA correction to raw measured isotopic ratios can be up to seventy permil or more. Therefore, not applying the correction, or misapplication of it could lead to significant inaccuracies in published isotope ratio data. Examples and ramifications of the latter are described in addition to a straightforward procedure for QSA under‐counting correction.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Jones, Clive (ORCID:0000000338247335), Fike, David. 2024-07-21. SIMS Data Correction Procedure for Quasi‐Simultaneous Arrival ( QSA ) Under‐counting and Ramifications of Misapplication. https://doi.org/10.1111/ggr.12570