Abstract
The LBNL team developed a multiport cell allowing for fluid flow, with electrical contacts for sample bias, and with reference electrodes. The cell is currently installed and operating in beam-line 11.0.2 of the ALS, the Berkeley Synchrotron. The cell is closed by a membrane consisting of several stacked layers of graphene (1 to 3). The graphene membrane is supported on a 100 nm thick Si3N4 membrane covered with a 20 nm thin gold film for improved adhesion of the graphene and to ensure electrical conductivity. The Si3N4 film is perforated with a periodic array of holes 1 to 2 micrometers diameter. The cell can withhold an internal pressure (gas or liquid) of up to 10 atmospheres for in situ/operando studies using X-ray Photoelectron emission (XPS), and X-ray Absorption Spectroscopy (XAS). It is also ideal for tip-enhanced nano-Infrared Spectroscopy studies of molecules at the interface between the graphene and the liquid or gas filling the cell. Materials studied in the LBNL laboratory are in the form of nanoparticles, or thin films (few nm) deposited on the side of the graphene membrane facing the solution or gas inside the cell.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Salmeron, Miquel [Lawrence Berkeley National Laboratory (LBNL), Berkeley, CA (United States)] (ORCID:0000000228878128), Alsem, Daan Hein [Hummingbird Precision Machine Company, Lacey, WA (United States)], Salmon, Norman [Hummingbird Precision Machine Company, Lacey, WA (United States)]. 2024-12-04. Multiport high-pressure synchrotron x-ray microscopy cell (CRADA Final Report). https://doi.org/10.2172/2588135
Cite the original work for its findings. Save a collection to share your selection of sources.