BIOASSAY VESSEL FAILURE ANALYSIS
Abstract
Two high-pressure bioassay vessels failed at the Savannah River Site during a microwave heating process for biosample testing. Improper installation of the thermal shield in the first failure caused the vessel to burst during microwave heating. The second vessel failure is attributed to overpressurization during a test run. Vessel failure appeared to initiate in the mold parting line, the thinnest cross-section of the octagonal vessel. No material flaws were found in the vessel that would impair its structural performance. Content weight should be minimized to reduce operating temperature and pressure. Outer vessel life is dependent on actual temperature exposure. Since thermal aging of the vessels can be detrimental to their performance, it was recommended that the vessels be used for a limited number of cycles to be determined by additional testing.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Vormelker, P. 2008-09-22. BIOASSAY VESSEL FAILURE ANALYSIS. https://searchnasa.com/record/osti%3A941107/bioassay-vessel-failure-analysis