Abstract
Pacific Northwest National Laboratory (PNNL) has demonstrated significant advancement in using beta-gamma coincidence detectors to detect a wide range of radioxenon isotopes. To obtain accurate activities with the detector it must be properly calibrated by measuring a series of calibration gas samples. The data is analyzed to create the calibration block used in the International Monitoring System file format. Doing the calibration manually has proven to be tedious and prone to errors, requiring a high degree of expertise. The Beowulf graphical user interface (GUI) is a software application that encompasses several components of the calibration task and generates a calibration block, as well as, a detailed report describing the specific calibration process used. This additional document can be used as a Quality assurance certificate to assist in auditing the calibration. This paper consists of two sections. Section 1 will describe the capabilities of Beowulf and section 2 will be a representative report generated or the 137Cs calibration and quality assurance source.