Characterization of Diffraction-Enhanced Imaging

Abstract

Diffraction-enhanced imaging (DEI) is a new x-ray imaging modality that has been shown to enhance contrast between normal and cancerous breast tissues. In this study, diffraction-enhanced imaging in computed tomography (DEI-CT) mode was used to quantitatively characterize the refraction contrasts of the organized structures associated with invasive human breast cancer. Using a high-sensitivity Si (3 3 3) reflection, the individual features of breast cancer, including masses, calcifications and spiculations, were observed. DEI-CT yields 14, 5 and 7 times higher CT numbers and 10, 9 and 6 times higher signal-to-noise ratios (SNR) for masses, calcifications and spiculations, respectively, as compared to conventional CT of the same specimen performed using the same detector, x-ray energy and dose. Furthermore, DEI-CT at ten times lower dose yields better SNR than conventional CT. In light of the recent development of a compact DEI prototype using an x-ray tube as its source, these results, acquired at a clinically relevant x-ray energy for which a pre-clinical DEI prototype currently exists, suggest the potential of clinical implementation of mammography with DEI-CT to provide high-contrast, high-resolution images of breast cancer (Parham 2006 PhD Dissertation University of North Carolina at Chapel Hill).

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Kao, T, Connor, D, Dilmanian, F, Faulconer, L, Liu, T, Parham, C, Pisano, E, Zhong, Z. 2009-01-01. Characterization of Diffraction-Enhanced Imaging. https://doi.org/10.1088/0031-9155%2F54%2F10%2F019

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related reports

High Multiplicity Trigger for long-lived particles in CMS detector

Searches for long-lived particles (LLPs) at the CMS experiment often involve unconventional event topologies that are difficult to efficiently select using standard trigger strategies. To improve sensitivity to such signatures during LHC Run 3 operation, a dedicated High Multiplicity Trigger (HMT) has been developed and deployed in the CMS trigger system. The trigger targets events containing unusually large numbers of hits in the CMS cathode strip chamber (CSC) muon detectors, a characteristic signature of several LLP scenarios involving displaced decays in the muon system. The HMT implementation, trigger logic, rate dependence with pileup, and operational stability are described. Optimized hit multiplicity thresholds are used to maintain acceptable trigger rates under high-luminosity and high-pileup conditions while preserving high efficiency across a broad range of LLP lifetimes and kinematic regimes. The trigger performance is evaluated using both simulated event samples and proton-proton collision data collected during Run 3 of the LHC. The HMT substantially extends the CMS sensitivity to non-standard signatures associated with LLP decays and provides a flexible platform for future searches for physics beyond the Standard Model.

47 OTHER INSTRUMENTATION↗