Testing Evaluation of the Electrochemical Organic Content Analyzer
The breadboard electrochemical organic content analyzer was evalauted for aerospace applications. An awareness of the disadvantages of expendables in some systems resulted in an effort to investigate ways of reducing the consumption of the analyzer's electrolyte from the rate of 5.17 kg/30 days. It was found that the electrochemical organic content analyzer can result in an organic monitor in the water quality monitor having a range of 0.1 to 100 mg/1 total organic carbon for a large number of common organic solutes. In a flight version it is anticipated the analyzer would occupy .0002 cu m, weigh 1.4 kg, and require 10 W or less of power. With the optimum method of injecting electrolyte into the sample (saturation of the sample with a salt) it would expend only 0.04 kg of electrolyte during 30 days of continuous operation.