Mass Spectra Deconvolution of Gaseous Mixtures Containing Volatile Organic Compounds
The Spacecraft Atmosphere Monitor (S.A.M.) analyzes all gaseous pollutants in the 1-150Th mass-to-charge range based on their positive-ion mass spectra. Of particular interest is the Trace Gas Analysis (TGA) mode of operation in which S.A.M detects minute amounts of volatile organic compounds (VOC). Pollutant component of the ambient air is sampled on demand in twenty full mass spectra per second. All spectra is accumulated in two-second intervals by an on-board Field Programmable Gate Arrays (FPGA) unit. Mass spectra represents the number of detected fragment ions in a given mass-to-charge ratio channel. Electron impact ionization of neutral molecules results in creation of numerous fragment ions and most of organic compounds will contribute several identical fragments. Despite the presence of these molecular isobars, we developed the deconvolution algorithm capable of identifying target species based on their characteristic fragmentation patterns. We investigate the efficiency of deconvolution algorithm as a function of mass resolution with which mass spectrum is acquired. Higher the mass resolution, better the chances are to distinguish between stereoisomers using their fragmentation propensities, but also results in higher data volumes to be processed by a set of small footprint software stacks hosted by an on-board computer. Finding the balance between deconvolution accuracy and generated data volume under time constrains and limited computing resources is the main topic of this study.