International Space Station Major Constituent Analyzer On-Orbit Performance
The Major Constituent Analyzer is a mass spectrometer based system that measures the major atmospheric constituents on the International Space Station. A number of limited-life components require periodic changeout, including the analyzer (ORU 02) and the verification gas assembly (ORU 08). The longest lasting ORU 02 was recently replaced after a record service length of 1033 days. The comparatively high performance duration may be attributable to a reduced inlet flow rate into the analyzer, resulting in increased ion pump lifetime; however, there may be other factors as well. A recent schedule slip for delivery of replacement verification gas led to a demonstration that the calibration interval could be extended on a short-term basis. An analysis of ORU 08 performance characteristics indicates that it is possible to temporarily extend the calibration interval from 6 weeks to 12 weeks if necessary.