The radiance and irradiance sensitivities of the TOMS are tabulated in table 1. The results obtained with FEL standard F-281 are actual while those computed for lamps F-239 and F-207 are for reference purposes as those standards were out of calibration at the time of use. The irradiance results were computed for the serial number 5 carousel surfaces, the flight unit, and the serial number 4 carousel surfaces, the flight backup unit. The data and results are presented tabularly and graphically. The computed instrument sensitivities based on the radiance results are higher than those resulting from the irradiance data (see figure's 9 a,b,&c). No quantitative explanation for this difference has been determined. The radiance data was obtained with the instrument in the NADIR detent while the irradiance data was obtained with the scan mirror rotated 90 degrees from NADIR to view the diffuser. The sensitivity plots show the pronounce effects of the instruments measured spectral bandwidth used in the computations. The radiance data, figures 5 a,b & c, are quite smooth. Sources of error in the calibrations include the normalized radiance, source lamp irradiance , diffuser BRDF, and geometry factors. These are discussed in the errors of calibration section included in this report.