Deuterated scintillators, such as deuterated xylene (C8D10) have been theorized to offer better pulse shape discrimination (PSD) and neutron spectrometry performance than traditional protiated scintillators due to the anisotropy of neutron scattering on deuterium. However, the light output responses of said deuterated scintillators to monoenergetic has not been characterized in its entirety at both high and low energies. Several time-of-flight experiments were conducted at the Idaho national Laboratory to characterize a pair of EJ-301D deuterated xylene detectors. The low energy detector response to quasi-monoenergetic neutrons was measured using a Cf-252 source, and the high energy detector response was measured using a D-T source and angles of 20, 45, and 60 deg, corresponding to theoretical neutron energies of 13.2, 10.4, and 8.2 MeV. During post-processing, the neutron and gamma bands were clearly able to be separated, even at low light outputs, solidifying the superior PSD capability of deuterated xylene. An exponential of fit of LO=0.63E-2.22*(1-exp(-0.25E)), which has good agreement with previously measured data, was found as the light output response from deuteron recoils. Additionally, measurements of AmBe, D-T, and D-D sources were unfolded using the light output response curve, which resulted in favorable agreement with theoretical values. Further testing of the response matrix by unfolding more sources (such as PuBe) needs to be done, but the unfolding accuracy for currently available sources suggests that deuterated xylene can be used to measure and unfold both fission, monoenergetic, and continuous spectra.