A Monolithic Spatial Heterodyne Raman Spectrometer: Initial Tests
Explore the source record for details and available documents.
SEARCH · Search NASA
Search indexed NASA NTRS and DOE OSTI research on propulsion, heat transfer, battery materials and energy systems. Follow report and document links to the original sources.
Quote a phrase for an exact phrase match. Source license links do not imply unrestricted reuse.
Explore the source record for details and available documents.
Explore the source record for details and available documents.
The availability in the last decade of improved Raman instrumentation using small, stable, intense lasers, sensitive CCD array detectors, and advanced fast grating systems enabled us to develop the Mars Microbeam Raman Spectrometer (MMRS), a field-portable Raman spectrometer with precision and accuracy capable of identifying minerals and their different compositions. For example, we can determine Mg cation ratios in pyroxenes and olivines to +/-0.1 on the basis of Raman peak positions. Feldspar is another major mineral formed in igneous systems whose characterization is important for determining rock petrogenesis and alteration. From their Raman spectral pattern, feldspars can be readily distinguished from ortho- and chain-silicates and from other tecto-silicates such as quartz and zeolites. We show here how well Raman spectral analysis can distinguish among members within the feldspar group.
A modified coronagraph has been proposed as a prototype of improved notch filters in Raman spectrometers. Coronagraphic notch filters could offer alternatives to both (1) the large and expensive double or triple monochromators in older Raman spectrometers and (2) holographic notch filters, which are less expensive but are subject to environmental degradation as well as to limitations of geometry and spectral range. Measurement of a Raman spectrum is an exercise in measuring and resolving faint spectral lines close to a bright peak: In Raman spectroscopy, a monochromatic beam of light (the pump beam) excites a sample of material that one seeks to analyze. The pump beam generates a small flux of scattered light at wavelengths slightly greater than that of the pump beam. The shift in wavelength of the scattered light from the pump wavelength is known in the art as the Stokes shift. Typically, the flux of scattered light is of the order of 10 7 that of the pump beam and the Stokes shift lies in the wave-number range of 100 to 3,000 cm 1. A notch filter can be used to suppress the pump-beam spectral peak while passing the nearby faint Raman spectral lines. The basic principles of design and operation of a coronagraph offer an opportunity for engineering the spectral transmittance of the optics in a Raman spectrometer. A classical coronagraph may be understood as two imaging systems placed end to end, such that the first system forms an intermediate real image of a nominally infinitely distant object and the second system forms a final real image of the intermediate real image. If the light incident on the first telescope is collimated, then the intermediate image is a point-spread function (PSF). If an appropriately tailored occulting spot (e.g., a Gaussian-apodized spot with maximum absorption on axis) is placed on the intermediate image plane, then the instrument inhibits transmission of light from an on-axis source. However, the PSFs of off-axis light sources are formed off axis - that is, away from the occulting spot - so that they become refocused onto the final image plane.
Interest in the use of Raman spectrometers has seen an increase in the fields of geology and planetary sciences due to the non-destructive insight Raman spectra may provide into the molecular makeup of a given sample. Advancements in Raman spectrometer hardware have allowed for compact instruments to have deployment capabilities directly on interplanetary missions, flexible usage conditions requiring no sample collection/preparation, and no need for daylight radiation shielding. As the amount of science which can be collected from a Raman spectrometer in a given amount of time increases, a bottleneck will be created in data analysis which leaves a need for a faster method of spectral data classification. Recent studies have shown that machine learning models are able to solve this problem by achieving high-accuracy classification. Liu et al4 found the convolutional neural network (CNN) held the highest classification accuracy (96% top 5) for single sample Raman data.
The Mars Surveyor missions that will be launched in April of 2001 will include a highly capable rover that is a successor to the Mars Pathfinder mission's Sojourner rover. The design goals for this rover are a total traverse distance of at least 10 km and a total lifetime of at least one Earth year. The rover's job will be to explore a site in Mars' ancient terrain, searching for materials likely to preserve a record of ancient martian water, climate, and possibly biology. The rover will collect rock and soil samples, and will store them for return to Earth by a subsequent Mars Surveyor mission in 2005. The Athena Mars rover science payload is the suite of scientific instruments and sample collection tools that will be used to perform this job. The specific science objectives that NASA has identified for the '01 rover payload are to: (1) Provide color stereo imaging of martian surface environments, and remotely-sensed point discrimination of mineralogical composition. (2) Determine the elemental and mineralogical composition of martian surface materials. (3) Determine the fine-scale textural properties of these materials. (4) Collect and store samples. The Athena payload has been designed to meet these objectives. The focus of the design is on field operations: making sure the rover can locate, characterize, and collect scientifically important samples in a dusty, dirty, real-world environment. The topography, morphology, and mineralogy of the scene around the rover will be revealed by Pancam/Mini-TES, an integrated imager and IR spectrometer. Pancam views the surface around the rover in stereo and color. It uses two high-resolution cameras that are identical in most respects to the rover's navigation cameras. The detectors are low-power, low-mass active pixel sensors with on-chip 12-bit analog-to-digital conversion. Filters provide 8-12 color spectral bandpasses over the spectral region from 0.4 to 1.1 micron Narrow-angle optics provide an angular resolution of 0.28 mrad/pixel, nearly a factor of four higher than that of the Mars Pathfinder and Mars Surveyor '98 cameras. Image compression will be performed using a wavelet compression algorithm. The Mini-Thermal Emission Spectrometer (Mini-TES) is a point spectrometer operating in -the thermal IR. It produces high spectral resolution (5 /cm) image cubes with a wavelength range of 5-40 gm, a nominal signal/noise ratio of 500:1, and a maximum angular resolution of 7 mrad (7 cm at a distance of 10 in). The wavelength region over which it operates samples the diagnostic fundamental absorption features of rockforming minerals, and also provides some capability to see through dust coatings that could tend to obscure spectral features. The mineralogical information that Mini-TES provides will be used to select from a distance the rocks and soils that will be investigated in more detail and ultimately sampled. Mini-TES is derived from the MO/MGS TES instrument, but is significantly smaller and simpler. The instrument uses an 8-cm Cassegrain telescope, a Michelson interferometer, and uncooled pyroelectric detectors. Along with its mineralogical capabilities, Mini-TES can provide information on the thermophysical properties of rocks and soils. Viewing upward, it can also provide temperature profiles through the martian atmospheric boundary layer. Elemental and Mineralogical Composition: Once promising samples have been identified from a distance using Pancam/Mini-TES, they will be studied in detail using up to three compositional sensors that can be placed directly against them by an Instrument Arm. The two compositional sensors, presently on the payload are an Alpha-Proton-X-Ray Spectrometer (APXS), and a Mossbauer Spectrometer. The APXS is derived closely from the instrument that flew on Mars Pathfinder. Radioactive alpha sources and three detection modes (alpha, proton, and x-ray) provide elemental abundances of rocks and soils to complement and constrain mineralogical data. The Athena APXS will have a revised mechanical design that will cut down significantly on backscattering of alpha particles from martian atmospheric carbon. It will also include a target of known elemental composition that will be used for calibration purposes. The Athena Mossbauer Spectrometer is a diagnostic instrument for the mineralogy and oxidation state of Fe-bearing phases, which are particularly important on Mars. The instrument measures the resonant absorption of gamma rays produced by a Co-57 source to determine splitting of nuclear energy levels in Fe atoms that is related to the electronic environment surrounding them. It has been under development for space flight for many years at the Technical University of Darmstadt. The Mossbauer Spectrometer (and the other arm instruments) will be able to view a small permanent magnet array that will attract magnetic particles in the martian soil. The payload may also include a Raman Spectrometer. If included, the Raman Spectrometer will provide precise identification of major and minor mineral phases. It requires no sample preparation, and is also sensitive to organics. Fine-Scale Texture: The Instrument Arm a also carries a Microscopic Imager that will obtain high-resolution monochromatic images of the same materials for which compositional data will be obtained. Its spatial resolution is 20 micron/pixel over a 1 cm depth of field, and 40 micron/pixel over a 1-cm depth of field. Like Pancam, it uses the same active pixel sensor detectors and electronics as the rover's navigation cameras. The Instrument Arm is a three degree-of-freedom arm that uses designs and components from the Mars Pathfinder and Mars Surveyor '98 projects. Its primary function is instrument positioning. Along with the instruments noted above, it also carries a brush that can be used to remove dust and other loose coatings from rocks. Sample Collection and Storage: Martian rock and soil samples will be collected using a low-power rotary coring drill called the Mini-Corer. An important characteristic of this device is that it can obtain intact samples of rock from up to 5 cm within strong boulders and bedrock, Nominal core dimensions are 8xl7 mm. The Mini-Corer drills a core to the commanded depth in a rock, shears it off, retains it, and extracts it. It can also acquire samples of loose soil, using soil sample cups that are pressed downward into loose material. The Mini-Corer can drill at angles from vertical to 45' off vertical. It has six interchangeable bits for long life. Mechanical damage to the sample during drilling is minimal, and heating is negligible. After acquisition, the sample may be viewed by the arm instruments, and/or placed in one of 104 compartments in the Sample Container. A subset of the acquired samples may be replaced with other samples obtained later if desired. The Sample Container has no moving parts, and is mounted external to the rover for easy removal by the Mars Surveyor 2005 flight system. Operation of the rover will make extensive use of automated onboard navigation and hazard avoidance capabilities. Otherwise, use of onboard autonomy is minimal. Data downlink capability is about 40 Mbit/sol, and the use of the Mars Surveyor '01 orbiter for data relay imposes a limit of at most two command cycles per sol. Because of the significant amount of time available between command cycles, all payload elements will be operated sequentially, rather than in parallel.; this approach also significantly simplifies operations and minimizes peak power usage. The landing site for the '01 rover has not been selected yet. Site selection will make as full use as possible of Mars Global Surveyor data, and will involve substantial input from the broad Mars science community. Summary: The following table describes the mass, power, providers, and key scientific objectives of all the major elements of the Athena payload. Additional Athena payload information may be found at: http://astrosun.tn.cornell.edu/athena/index.html. Additional information contained in the original.
The in Situ Spectroscopic Europa Explorer (iSEE) instrument is an ultra-compact laser-enabled Raman spectrometer instrument that meets the top-level science requirements for multiple future planetary in situ missions to explore the surface and atmospheric chemistry of planetary bodies across the Solar System. Enceladus, Europa, the Moon, Mars, and Venus are some of the primary targets for future NASA missions to search for extraterrestrial life and potentially habitable environments beyond Earth, further our understanding of the timing and formation of the Solar System and identify potentially viable economic resources such as water and/or valuable metal assets. We report on the advancement and space flight qualification of a compact, robust, solid-state laser operating at 515 nm that serves as the excitation source for the iSEE investigation. The iSEE instrument is being developed under a NASA Maturation of Instruments for Solar System Exploration (MatISSE) program. The iSEE laser is a diode pumped 1030 nm Yb:YAG microchip laser with a second harmonic generator to achieve an output wavelength of 515 nm and coupled to a multi-mode fiber for delivery to the Raman spectrometer probe. The fundamental 1030 nm laser operates at a pulse repetition frequency (PRF) of 1-10 kHz with pulse energy of 95 μJ and ~800 ps pulse width. A 15 mm long Type II KTP crystal is used for second harmonic generation from 1030 nm to 515 nm. The residual fundamental 1030 nm is separated from the 515 nm beam using a pair of dichroic filters. A focusing lens couples the 515 nm beam to the optical fiber to produce an output laser pulse energy of 20 μJ. A custom vacuum fiber feedthrough assembly was designed with two 105 μm core fibers, and two 200 μm core fibers. One fiber is needed for coupling the laser to the iSEE instrument, and the additionally fibers provide redundancy and design flexibility. The laser enclosure is pressurized with >1 atm of clean dry air. The laser has been designed to minimize the size, weight, and power (SWaP) for a lander instrument with minimal resources. The laser will go through environmental testing including vibration and thermal vacuum testing for space flight qualification. Here we discuss the design, trade studies, performance, and environmental qualificationtesting of the iSEE laser.
The sharp, nonoverlapping Raman bands for plagioclase, pyroxene, and olivine would be advantageous for on-surface, active mineralogical analysis of lunar materials. A robust, light-weight, low-power, rover-based Raman spectrometer with a laser exciting source, entirely transmission-mode holographic optics, and a charge-coupled device (CCD) detector could fit within a less than 20 cm cube. A sensor head on the end of an optical fiber bundle that carried the laser beam and returned the scattered radiation could be placed against surfaces at any desired angle by a deployment mechanism; otherwise, the instrument would need no moving parts. A modem micro-Raman spectrometer with its beam broadened (to expand the spot to 50-micrometer diameter) and set for low resolution (7/cm in the 100-1400/cm region relative to 514.5-nm excitation), was used to simulate the spectra anticipated from a rover instrument. We present spectra for lunar mineral grains, less than 1 mm soil fines, breccia fragments, and glasses. From frequencies of olivine peaks, we derived sufficiently precise forsterite contents to correlate the analyzed grains to known rock types and we obtained appropriate forsterite contents from weak signals above background in soil fines and breccias. Peak positions of pyroxenes were sufficiently well determined to distinguish among orthorhombic, monoclinic, and triclinic (pyroxenoid) structures; additional information can be obtained from pyroxene spectra, but requires further laboratory calibration. Plagioclase provided sharp peaks in soil fines and most breccias even when the glass content was high.
The chemical composition of Venus’ clouds has attracted considerable attention since the first atmospheric probes entered the planet in the early 1980s. We know from these early studies that cloud aerosols consist of micron and submicron droplets of sulfuric acid (~85%) and water (~15%), but outstanding questions still exist regarding their trace chemical composition, especially the identity of the unknown UV absorber(s) and the possible presence of organic material (Spacek & Benner, 2021). Veritas and DaVinci will launch at the end of the decade to advance our understanding of Venus’ surface and atmosphere, however these missions are not designed to sample and analyze cloud aerosols at trace levels. Raman spectroscopy is an ideal candidate for characterizing aerosols and has a track record in studies of Earth’s atmosphere. Raman provides broad chemical screening of organic and inorganic molecules within seconds, and Raman instrumentation can be miniaturized to meet the stringent size, weight, and power (SWaP) constraints of an atmospheric probe while retaining sensitivity and specificity that allow spectral fingerprinting of compounds and functional groups. Raman alone is insufficient to characterize trace constituents however, as limits of detection are typically in the 10s to 100s of ppm at best for planetary Raman spectrometers. Surface-Enhanced Raman Spectroscopy (SERS), a method that utilizes nanoscale materials interacting with a sample to enhance weak Raman signals by orders of magnitude when probed by a laser, has the potential to drastically improve the sensitivity of a Raman instrument while adding little-to-no SWaP nor complexity to a scientific payload. We aim to de-risk SERS technology for rapid infusion in future planetary science missions by developing custom SERS substrates tailored towards Venus atmospheric exploration. We have developed SERS substrates consisting of nanotextured silver films and tested them on potential Venus-relevant organic compounds, with preliminary results indicating sub-ppm sensitivity. Additionally, we have planned tests to investigate durability, longevity, and other key characteristics that impact SERS substrate performance to prepare SERS technology for flight.
Mass and Raman spectrometers represent progressive analytical platforms for future in situ lander missions to explore the surface chemistry of planetary bodies across the Solar System. Europa, the Moon, and Mars are some of the primary targets for future NASA missions to search for extraterrestrial life and potentially habitable environments beyond Earth, further our understanding of the timing and formation of the Solar System and identify potentially viable economic resources such as water and/or valuable metal assets. The CORALS (Characterization of Ocean Residues And Life Signatures) and CRATER (Characterization of Regolith And Trace Economic Resources) instruments are UV laser-enabled OrbitrapTM mass spectrometers currently under development for prospective lander missions to Europa and the Moon, respectively. The Raman Mass Spectrometer (RAMS) is a dual wavelength (UV and visible) laser enabled hybrid instrument that incorporates laser desorption/ ionization mass spectrometry (LDMS) and micro-Raman microspectroscopy imaging (μRS) into a compact instrument package for multiple planetary missions to the inner and outer Solar System. The Pulse Laser Ablation Sampling and Mass Analysis (PLASMA) instrument combines a multi-wavelength laser (deep-UV to near-IR), plasma source, and collision cell with a heritage quadrupole mass spectrometer (QMS) to enable high precision trace element analyses for a future Lunar lander instrument. We report on the advancement of four compact, robust, and high technology readiness level (TRL) solid state lasers operating at wavelengths 213 nm, 266 nm, 515 nm, 532 nm and 1064 nm that serve as the sampling and ionization sources, for the CRATER, CORALS, RAMS and PLASMA investigations.
Given that, in most instances, minimal sample preparation is required and due to its contactless instrument design, Raman spectroscopy is one of the most versatile vibrational spectroscopic techniques for the chemical analysis of environmental and biological specimens. The diversity of applications of Raman spectroscopy ranges anywhere from art [1] to planetary science missions [2]. The advancement in the use of Raman spectroscopy in Solar System missions, notably in post-mission sample return analysis, requires a spectral library holding the broad range of specimens that could be found in Solar System sources. For this purpose, we have initiated the development of a Raman spectral database (Ramdb) at NASA Ames Research Center. Currently, the database includes experimental and theoretical Raman spectra of PAHs [3, 4], as well as laboratory Raman spectra of amino acids, carbon allotropes, minerals, and analogs relevance to Earth Sciences [5], Exobiology [6], Planetary [7], and Astrochemistry [8] to name just a few examples. Ramdb can be found on the web at www.astrochemistry.org/ramdb, where raw and processed Raman spectra can be downloaded in CSV format. The laboratory Raman spectra are measured using a laser Raman spectrometer (JASCO NRS-5500-532QRI). The Raman instrument is equipped with three excitation lasers, with wavelengths of 405, 532, and 785 nm. A clean silicon substrate is used as the internal standard for wavenumber calibration. Powdered samples were prepared (microscopic >10 um, grounded microscopic < 10 um) on glass slides. Some raw data exhibited a background signal arising as a combination of laser-induced fluorescence from the sample. To correct this background, we developed a Python pipeline that uses open-source Python libraries. Ramdb provides both raw and processed (using Python pipeline) data, which includes tabulated Raman shift transitions and other measurement details. The theoretical Raman band positions of PAHs (pyrene monomers and tetramer clusters) were computed using density functional theory (DFT) with the help of the Gaussian 16 suite of programs [9]. In the near future, Ramdb will serve as a repository of Raman spectral data from Laboratory Astrophysics and Planetary Science experiments involving the irradiation of organic compounds under simulated space and planetary conditions. In addition, online and offline tools will be developed for utilising the database for comparison to the user’s sample.
A modification to a commercial FT-Raman spectrometer is presented for the elimination of thermal backgrounds in FT-Raman spectra. The modification involves the use of a mechanical chopper to modulate the CW laser, remote collection of the signal via fiber optics, and connection of a dual-phase digital signal processor lock-in amplifier between the detector and the spectrometer's collection electronics to demodulate and filter the optical signals. The resulting modulated FT-Raman fiber-optic spectrometer is capable of completely eliminating thermal backgrounds at temperatures exceeding 370 C. In addition, the signal/noise of generated Raman spectra is greater than for spectra collected with the conventional FT-Raman under identical conditions and incident laser power. This is true for both room-temperature and hot samples. The method allows collection of data using preexisting spectrometer software. The total cost of the modification (excluding fiber optics) is approximately $3000 and requires less than 2 h to implement. This is the first report of Fr-Raman spectra collected at temperatures in excess of 300 C in the absence of thermal backgrounds.
A modification to a commercial Fourier Transform (FT) Raman spectrometer is presented for the elimination of thermal backgrounds in the FT Raman spectra. The modification involves the use of a mechanical optical chopper to modulate the continuous wave laser, remote collection of the signal via fiber optics, and connection of a dual-phase digital-signal-processor (DSP) lock-in amplifier between the detector and the spectrometer's collection electronics to demodulate and filter the optical signals. The resulting Modulated Fourier Transform Raman Fiber-Optic Spectrometer is capable of completely eliminating thermal backgrounds at temperatures exceeding 300 C.
The in Situ Spectroscopic Europa Explorer (iSEE) instrument is an ultra-compact laser-enabled Raman spectrometer instrument that meets the top-level science requirements for multiple future planetary in situ missions to explore the surface and atmospheric chemistry of planetary bodies across the Solar System. Enceladus, Europa, the Moon, Mars, and Venus are some of the primary targets for future NASA missions to search for extraterrestrial life and potentially habitable environments beyond Earth, further our understanding of the timing and formation of the Solar System, and identify potentially viable economic resources such as water and/or valuable metal assets. We report on the advancement and space flight qualification of a compact, robust, solid-state laser operating at 515 nm that serves as the excitation source for the iSEE investigation.
A multiplex coherent Raman spectrometer (10) and spectroscopy method rapidly detects and identifies individual components of a chemical mixture separated by a separation technique, such as gas chromatography. The spectrometer (10) and method accurately identify a variety of compounds because they produce the entire gas phase vibrational Raman spectrum of the unknown gas. This is accomplished by tilting a Raman cell (20) to produce a high-intensity, backward-stimulated, coherent Raman beam of 683 nm, which drives a degenerate optical parametric oscillator (28) to produce a broadband beam of 1100-1700 nm covering a range of more than 3000 wavenumber. This broadband beam is combined with a narrowband beam of 532 nm having a bandwidth of 0.003 wavenumbers and focused into a heated windowless cell (38) that receives gases separated by a gas chromatograph (40). The Raman radiation scattered from these gases is filtered and sent to a monochromator (50) with multichannel detection.
An instrumentation system that would include an ultraviolet (UV) laser or light-emitting diode, an infrared (IR) laser, and the equivalent of an IR Raman spectrometer has been proposed to enable noncontact identification of hazardous biological agents and chemicals. In prior research, IR Raman scattering had shown promise as a means of such identification, except that the Raman-scattered light was often found to be too weak to be detected or to enable unambiguous identification in practical applications. The proposed system would utilize UV illumination as part of a two-level optical-pumping scheme to intensify the Raman signal sufficiently to enable positive identification.
Fe-Ti-Cr-Oxide minerals contain much information about rock petrogenesis and alteration. Among the most important in the petrology of common intrusive and extrusive rocks are those of the FeO-TiO2-Cr2O3 compositional system chromite, ulv spinel-magnetite, and ilmenite-hematite. These minerals retain memories of oxygen fugacity. Their exsolution into companion mineral pairs give constraints on formation temperature and cooling rate. Laser Raman spectroscopy is anticipated to be a powerful technique for characterization of materials on the surface of Mars. A Mars Microbeam Raman Spectrometer (MMRS) is under development. It combines a micro sized laser beam and an automatic point-counting mechanism, and so can detect minor minerals or weak Raman-scattering phases such as Fe- Ti-Cr-oxides in mixtures (rocks & soils), and provide information on grain size and mineral mode. Most Fe-Ti-Cr-oxides produce weaker Raman signals than those from oxyanionic minerals, e.g. carbonates, sulfates, phosphates, and silicates, partly because most of them are intrinsically weaker Raman scatters, and partly because their dark colors limit the penetration depth of the excitation laser beam (visible wavelength) and of the Raman radiation produced. The purpose of this study is to show how well the Fe-Ti-Cr-oxides can be characterized by on-surface planetary exploration using Raman spectroscopy. We studied the basic Raman features of common examples of these minerals using well-characterized individual mineral grains. The knowledge gained was then used to study the Fe-Ti-Cr-oxides in Martian meteorite EETA79001, especially effects of compositional and structural variations on their Raman features.
The vibrational spectrum of a high temperature (330 C) polymerization reaction was successfully monitored in real time using a modulated fiber-optic FT-Raman spectrometer. A phenylethynyl terminated monomer was cured, and spectral evidence for two different reaction products was acquired. The products are a conjugated polyene chain and a cyclized trimer. This is the first report describing the use of FT-Raman spectroscopy to monitor a high temperature (greater than 250 C) reaction in real time.