Miniaturized system of a gas chromatograph coupled with a Paul ion trap mass spectrometer

Miniature gas chromatography (GC) and miniature mass spectrometry (MS) instrumentation has been developed to identify and quantify the chemical compounds present in complex mixtures of gases. The design approach utilizes micro-GC components coupled with a Paul quadrupole ion trap (QIT) mass spectrometer. Inherent to the system are high sensitivity, good dynamic range, good QIT resolution, low GC flow-rates to minimize vacuum requirements and the need for consumables; and the use of a modular approach to adapt to volatile organic compounds dissolved in water or present in sediment. Measurements are reported on system response to gaseous species at concentrations varying over four orders of magnitude. The ability of the system to deal with complicated mixtures is demonstrated, and future improvements are discussed. The GC/QIT system described herein has a mass, volume and power that are, conservatively, one-twentieth of those of commercial off-the-shelf systems. Potential applications are to spacecraft cabin-air monitoring, robotic planetary exploration and trace-species detection for residual gas analysis and environmental monitoring.

Miniaturization/instrumentation/methods↗

Compact Two-step Laser Time-of-Flight Mass Spectrometer for in Situ Analyses of Aromatic Organics on Planetary Missions

RATIONALE A miniature time-of-flight mass spectrometer has been adapted to demonstrate two-step laser desorption-ionization (LOI) in a compact instrument package for enhanced organics detection. Two-step LDI decouples the desorption and ionization processes, relative to traditional laser ionization-desorption, in order to produce low-fragmentation conditions for complex organic analytes. Tuning UV ionization laser energy allowed control ofthe degree of fragmentation, which may enable better identification of constituent species. METHODS A reflectron time-of-flight mass spectrometer prototype measuring 20 cm in length was adapted to a two-laser configuration, with IR (1064 nm) desorption followed by UV (266 nm) postionization. A relatively low ion extraction voltage of 5 kV was applied at the sample inlet. Instrument capabilities and performance were demonstrated with analysis of a model polycyclic aromatic hydrocarbon, representing a class of compounds important to the fields of Earth and planetary science. RESULTS L2MS analysis of a model PAH standard, pyrene, has been demonstrated, including parent mass identification and the onset o(tunable fragmentation as a function of ionizing laser energy. Mass resolution m/llm = 380 at full width at half-maximum was achieved which is notable for gas-phase ionization of desorbed neutrals in a highly-compact mass analyzer. CONCLUSIONS Achieving two-step laser mass spectrometry (L2MS) in a highly-miniature instrument enables a powerful approach to the detection and characterization of aromatic organics in remote terrestrial and planetary applications. Tunable detection of parent and fragment ions with high mass resolution, diagnostic of molecular structure, is possible on such a compact L2MS instrument. Selectivity of L2MS against low-mass inorganic salt interferences is a key advantage when working with unprocessed, natural samples, and a mechanism for the observed selectivity is presented.

Getty, Stephanie↗

Sub-ppb Oxygen Contaminant Detection in Semi-Conductor Processing

Gaseous contaminants such as oxygen, water vapor, nitrogen and hydrocarbons are often present in the processing environment in semiconductor device fabrication and in containerless materials processing. The contaminants arise as a result of outgassing from hot surfaces or they may be part of the impurities in commercial ultra-high purity gases. Among these gaseous contaminants, oxygen is the most reactive and, therefore, has the most adverse effects on the end product. There has been an intense effort at the Jet Propulsion Laboratory to develop different types of oxygen sorbents to reduce oxygen concentration in a microgravity processing environment to sub-ppb (parts-per-billion) levels. Higher concentrations can lead to rapid surface oxide formation, hence reducing the quality of semiconductor devices. If the concentration of oxygen in a processing chamber at 1000oC is in the ppb level, it will only take approximately 10 seconds for an oxide layer to form on the surface of a sample. The interaction of oxygen with the water surface can lead to the formation of localized defects in semi-conductor devices, hence decreasing the manufacturing yield. For example, efficient production of 64 Mb RAM chips requires contaminations below ppb levels. This paper describes a technique for measuring trace quantities of oxygen contaminants by recording the monoatomic negative ions, O-, using mass spectrometry. The O- formation from the e--O2 interaction utilizes the electron dissociative attachment method that is greatly enhanced at the resonant energy (6.8 eV). The device combines a small gridded electron ionizer with a compact mass spectrometer. The concentrations of oxygen have been measured using the method of standard additions by diluting O2 in N2. The lowest detection limit obtained was 1.2 kHz (O- count rate) at a concentration of 10-10, corresponding to 0.1 ppb.

Sub-ppb Contaminant↗

Method for the Collection, Gravimetric and Chemical Analysis of Nonvolatile Residue (NVR) on Surfaces

Nonvolatile residue (NVR), sometimes referred to as molecular contamination is the term used for the total composition of the inorganic and high boiling point organic components in particulates and molecular films deposited on critical surfaces surrounding space structures, with the particulate and NVR contamination originating primarily from pre-launch operations. The "nonvolatile" suggestion from the terminology NVR implies that the collected residue will not experience much loss under ambient conditions. NVR has been shown to have a dramatic impact on the ability to perform optical measurements from platforms based in space. Such contaminants can be detected early by the controlled application of various detection techniques and contamination analyses. Contamination analyses are the techniques used to determine if materials, components, and subsystems can be expected to meet the performance requirements of a system. Of particular concern is the quantity of NVR contaminants that might be deposited on critical payload surfaces from these sources. Subsequent chemical analysis of the contaminant samples by infrared spectroscopy and gas chromatography mass spectrometry identifies the components, gives semi-quantitative estimates of contaminant thickness, indicates possible sources of the NVR, and provides guidance for effective cleanup procedures. In this report, a method for the collection and determination of the mass of NVR was generated by the authors at NASA Langley Research Center. This report describes the method developed and implemented for collecting NVR contaminants, and procedures for gravimetric and chemical analysis of the residue obtained. The result of this NVR analysis collaboration will help pave the way for Langley's ability to certify flight hardware outgassing requirements in support of flight projects such as Stratospheric Aerosol and Gas Experiment III (SAGE III), Clouds and the Earth's Radiant Energy System (CERES), Materials International Space Station Experiment - X (MISSE-X), and Doppler Aerosol Wind Lidar (DAWN).

Gordon, Keith↗

In situ analysis of planetary surfaces for organic materials.

Discussion of the possibility of investigating the physical, chemical, and biological state of other planets in the hope of encountering one or more that is presently in a state prior to or just after the evolution of living systems. It may then be possible to observe the chemistry at and near the surface, evaluate the sources of energy available, and search for the presence or absence of living systems, and their degree of similarity to terrestrial ones. The mass spectrometry technique is considered to be the most appropriate. A system consisting of a series of sample ovens which can be consecutively attached to the gas stream of a small gas chromatograph which is in turn connected to a mass spectrometer via a carrier gas separator has been developed for the exploration of the surface of Mars.

Biemann, K.↗

High-Resolution Spectroscopy of Mars: Recent Results and Implications for Atmospheric Evolution

It is believed that Earth, Venus, and Mars were formed by the same rocky and icy planetesimals, which resembled meteorites and comets in their composition, respectively. These planets are thus expected to have initially had the same chemical and isotope composition. Scaling the mass of the terrestrial ocean by the planetary mass ratio, the expected initial H2O abundance on Mars is a layer of about 1 km thick. Scaling the abundance of CO2 on Venus, the expected initial CO2 abundance on Mars is 15 bars. Evidently, significant parts of the initial H2O and CO2 abundances have been lost. Intense meteorite impact erosion and hydrodynamic escape of hydrogen (which could drag to escape more heavy species) were dominant loss processes in the first 0.8 Byr. Later, atmospheric sputtering by O+ ions resulted in the dissociation of CO2 and massive losses of O, C, and H. Formation of carbonates also reduced CO2 to its present abundance which currently exists in the atmosphere, on the polar caps, and is absorbed by regolith. Water loss is currently due to thermal escape of H and nonthermal escape of O, both formed by photodissociation of H2O. All loss processes resulted in fractionation of the H, O, and C isotopes. Therefore, the current isotope ratios in H2O and CO2 are clues to the history of volatiles on Mars. There are three tools to study H2O and CO2 isotopes in the martian atmosphere: (i) mass spectrometry from landing probes, (ii) analyses of Mars' gases trapped in the SNC meteorites which were ejected from Mars, and (iii) high-resolution spectroscopy of the H2O andCO2 bands. Method (i) is the best but is the most expensive. Mass spectrometers to be used should be designed for high-precision isotope measurements. Method (ii) makes it possible to reach an uncertainty +/- 0.1%. However, the obtained results are affected by some uncontrolled interactions: isotope fractionations of (1) trapped gases and (2) those released in pyrolysis, (3) contribution of the impactor, isotope exchanges (4) in the terrestrial environment and (5) with the host rock during pyrolysis. Therefore, the spectroscopic data are of great interest, though their formal accuracy is lower. High-resolution spectroscopy is also a tool to study the current atmosphere of Mars by mapping of some photochemically important species and searching for some minor constituents and their variations. Additional information is contained in the original extended abstract.

Krasnopolsky, V. A.↗

GC-MS in the analysis of organic compounds in meteorites

The application of an analytical approach combining gas chromatography with mass spectrometry (GC-MS) has shown that the amino acid composition of meteorite extracts is quite complex. A computer was used in the evaluation of the data obtained in the investigations. The computer programs developed have been concerned solely with the mass spectra of amino acids. Specialized programs have been written to determine the number of carbon atoms in an amino acid which is a member of any of three subclasses.

Lawless, J. G.↗

Knudsen measurements of the sublimation and the heat of formation of GeSe

Knudsen effusion studies of the sublimation of polycrystalline GeSe have been performed employing mass spectrometry. The results demonstrate that GeSe vaporizes congruently under present experimental conditions according to the reaction: GeSe(s) yields GeSe(g). The mean values for the third-law heat and second-law entropy of reaction based on direct mass-loss data are 42.0 + or - 1.5 kcal/mole and 42.3 + or - 1.6 eu respectively. From these data the standard heat of formation was calculated to be -10.1 + or - 2.0 kcal/mole, and the standard absolute entropy was determined to be 16.9 + or - 2.0 eu.

Wiedemeier, H.↗

A Quadrupole Ion Trap Mass Spectrometer for Quantitative Analysis of Nitrogen-Purged Compartments within the Space Shuttle

To enter orbit, the Space Shuttle burns 1.8 million liters of liquid hydrogen combined with 0.8 million liters of liquid oxygen through three rocket engines mounted in the aft. NASA monitors the nitrogen-purged aft compartment for increased levels of hydrogen or oxygen in order to detect and determine the severity of a cryogenic fuel leak. Current monitoring is accomplished with a group of mass spectrometer systems located as much as 400 feet away from the Shuttle. It can take up to 45 seconds for gas to reach the mass spectrometer, which precludes monitoring for leaks in the final moments before liftoff (the orbiter engines are started at T-00:06 seconds). To remedy the situation, NASA is developing a small rugged mass spectrometer to be used as point-sensors around the Space Shuttle. As part of this project, numerous mass analyzer technologies are being investigated. Presented here are the preliminary results for one such technology, quadrupole ion trap mass spectrometry (QITMS). A compact QITMS system has been developed in-house at the University of Florida for monitoring trace levels of four primary gases, hydrogen, helium, oxygen, and argon, all in a nitrogen background. Since commercially available QITMS systems are incapable of mass analysis at m/z(exp 2), the home-built system is preferred for the evaluation of QITMS technology.

Ottens, Andrew K.↗

Laser microprobe characterization of C species in Interplanetary Dust Particles (IDP)

Preliminary results of a study whose aim is the characterization of carbon (C) species in microvolumes of materials by means of laser ionization mass spectrometry (LIMS) are presented. The LIMS instrument employs a pulsed UV laser to produce nearly instantaneous vaporization and ionization of materials, followed by acceleration and time-of-flight analysis of the ions produced. LIMS provides a survey technique with nearly simultaneous acquisition of mass spectra covering the entire elemental range. The main limitation of the LIMS technique at present is its limited ability to perform quantitative analysis, due in part to insufficient knowledge of the mechanism of laser-solid interaction. However, considerable effort is now being directed at making LIMS a more quantitative technique. A variety of different C samples, both natural and man made were analyzed to establish the ability of LIMS to differentiate among the various C phases. The results of preliminary analyses performed on meteoritical and interplanetary dust samples are also presented. The C standards selected for the LIMS characterization range from essentially amorphous soot to diamond, which exhibits the highest degree of ordering.

Dibrozolo, F. R.↗

Water-soluble Organic Components in Aerosols Associated with Savanna Fires in Southern Africa: Identification, Evolution and Distribution

During the SAFARI 2000 field campaign, both smoke aerosols from savanna fires and haze aerosols in the boundary layer and in the free troposphere were collected from an aircraft in southern Africa. These aerosol samples were analyzed for their water-soluble chemical components, particularly the organic species. A novel technique, electrospray ionization-ion trap mass spectrometry, was used concurrently with an ion chromatography system to analyze for carbohydrate species. Seven carbohydrates, seven organic acids, five metallic elements, and three inorganic anions were identified and quantified. On the average, these 22 species comprised 36% and 27% of the total aerosol mass in haze and smoke aerosols, respectively. For the smoke aerosols, levoglucosan was the most abundant carbohydrate species, while gluconic acid was tentatively identified as the most abundant organic acid. The mass abundance and possible source of each class of identified species are discussed, along with their possible formation pathways. The combustion phase of a fire had an impact on the chemical composition of the emitted aerosols. Secondary formation of sulfate, nitrate, levoglucosan, and several organic acids occurred during the initial aging of smoke aerosols. It is likely that under certain conditions, some carbohydrate species in smoke aerosols, such as levoglucosan, were converted to organic acids during upward transport.

Gao, Song↗

Reaction of O2(+)(X 2Pi sub g) with H2, D2, and HD - Guided ion beam studies, MO correlations, and statistical theory calculations

Guided ion-beam mass spectrometry is used to examined the reactions of vibrationally cold ground-state O2(+)(X 2Pi sub g) with H2, D2, and HD. The energy dependence of the absolute integral cross sections from thermal energy to over 4 eV are measured in the center-of-mass frame of reference. Results are also presented for internally excited O2(+) ions reacting with D2 and HD. The results are consistent with the dominant state being the a 4Pi sub u electronic state. The experimental excitation functions are analyzed in detail and interpreted by extending the molecular orbital correlation arguments of Mahan (1971) and by comparison with results of statistical phase space theory and with a theory that predicts a tight transition state.

Weber, M. E.↗

Evaluation of Volatile Species in Green Monopropellant Project

NASA is interested in green monopropellants to replace hydrazine in reaction control systems (RCSs). Some current NASA programs require reduced vapor pressure and low toxicity monopropellant (green) and superior performance (specific impulse and density) formulations. Earlier vapor phase studies of a candidate green monopropellant at the NASA White Sands Test Facility (WSTF) showed the presence of a volatile species that warranted further investigation. The purpose of this study was to further characterize the volatile species and to evaluate it. The evaluation was with respect to whether the volatile species was an impurity or how it is formed, and to use that information to examine whether its presence as an impurity can be eliminated during formulation. The evaluation also considered whether formation of the volatile impurity could be prevented while not compromising the propellant. To reduce variables associated with evaluation of the propellant formulation as a whole, a precursor to one of the individual components in the propellant formulation was subjected to a NASA Standard 6001B Flammability, Off-gassing, and Compatibility Requirements and Test Procedures "Determination of Off-gassed Products (Test 7)". Testing took place in the NASA WSTF Molecular Desorption and Analysis Laboratory. One gram of the precursor was placed in a flask within a specimen container. After thermal conditioning for 72 +/- 1 h at 50 +/- 3 deg C (122 +/- 5 deg F), the atmosphere inside the specimen container was analyzed for off-gassed compounds by cryotrap gas chromatography-mass spectrometry (GC-MS) and fixed sample loop GC-flame ionization detection (GC-FID). The specimen container used was glass to minimize potential catalytic surfaces. The identification of compounds was difficult due to the complexity of the vapor phase concentrations and overlapping chromatographic peaks and mass spectra. However, eleven compounds were specifically identified and five compounds or classes of compounds were reported as unidentified. Quantitation of most of the compounds, including unidentified compounds, was as methane. Quantitating compounds or classes of compounds that were detected but for which specific calibration is not established as methane is in accordance with the Test 7 standard protocol. The thermal decomposition temperature of the precursor was significantly higher than the test temperature. Based on thermal decomposition temperature and on an examination of the structure and chemistry of the identified volatile species, the presence of the volatile species appears to be chemically reasonable with respect to the propellant formulation and is at this time attributed to impurities. Further examination of the overall propellant formulation process (including the individual components' synthesis processes) and process quality control (including purity of reagents and possible decomposition reactions) is indicated.

Greene, Benjamin↗

Dating Igneous Rocks Using the Potassium–Argon Laser Experiment (KArLE) Instrument: A Case Study for ~380 Ma Basaltic Rocks

We report new K–Ar isochron data for two ~380 Ma basaltic rocks, using an updated version of the Potassium–Argon Laser Experiment (KArLE), which is being developed for future in situ dating of planetary materials. These basalts have K contents comparable with those of lunar KREEP basalts or igneous lithologies found by Mars rovers, whereas previous proof‐of‐concept studies focused primarily on more K‐rich rocks. We aim to measure these analogous samples to show the advancing capability of in situ K–Ar geochronology. Combining laser‐induced breakdown spectroscopy (LIBS), mass spectrometry (MS), and microscopic analyses, we measured the abundance of K and 40Ar from 23 spots on the basalt samples. We then constructed K–Ar isochron plots from these rocks. The breadboard instrument consists of flight‐equivalent devices including a 30‐mJ Nd:YAG laser and a quadrupole mass spectrometer. Despite much lower K abundances than in previous studies, the isochron slopes yielded 380 ± 44 Ma and 398 ± 50 Ma for 380.7‐Ma and 373.5‐Ma rocks, respectively, indicating that accuracy better than 25 Ma (<7%) is achievable with our instrument. The isochron intercepts both yielded trapped 40Ar approximately 1 × 10−6 cm3 STP/g. Our experimental results demonstrate that accurate and precise measurements are possible using the KArLE approach on basaltic rocks, which are ubiquitous on planetary surfaces, and are useful in addressing a wide range of questions in planetary science.

Yuichiro Cho↗

Search for isotopic anomalies in oldhamite (CaS) from unequilibrated (E3) enstatite chondrites

The Ca isotopic compositions of 32 oldhamite (CaS) grains from the Qingzhen (EH3), MAC88136 (EL3), and Indarch (EH4) enstatite chondrites were determined by ion microprobe mass spectrometry. Also measured were the S isotopic compositions of eight oldhamite, two niningerite (MgS), and seven troilite (FeS) grains. The S isotopic compositions of all minerals are normal, but oldhamite grains of the first two meteorites exhibit apparent small Ca-48 excesses and deficits that are correlated with isotopic mass fractionation as determined from the Ca-40-Ca-44 pair. The interpretation of these results is complicated by the fact that none of the established mass fractionation laws can account for the data in the Norton County oldhamite standard. The method of analysis is carefully scrutinized for experimental artifacts. Neither interferences nor any known mass frationation effect can satisfactorily explain the observed small deviations from normal isotopic composition. If these are truly isotopic anomalies, they are much smaller than those observed in hibonite. The nucleosynthetic origin of Ca isotopes is discussed.

Lundberg, Laura L.↗

Electron-impact promoted fragmentation of alkyl-n-/1-phenylethyl/-carbamates of primary, secondary and tertiary alcohols.

The mass spectra of twenty alkyl carbamates derived from primary, secondary, and tertiary alcohols are investigated, using deuterium labeling and high-resolution mass spectrometry. Several of the primary and secondary alcohol derivatives are found to yield an ion formally equivalent to the product ion of a McLafferty rearrangement. Deuterium labeling, however, revealed a lack, in these carbamate derivatives, of the usual site specificity associated with the McLafferty rearrangement. A double hydrogen rearrangement process was observed in the mass spectra of several carbamates derived from tertiary alcohols.

Pereira, W. E.↗

Mass spectrometric investigation of the vaporization of sodium and potassium chromates: Preliminary results

Knudsen cell mass spectrometry was used to study the vaporization of sodium and potassium chromates. For both salts, the vaporization proceeds predominately by the reactions M2CrO4(c)=2M(g)+5/4O2(g)+1/2 Cr203(s) and M2CrO4(c)=M2CrO4(g) where M = Na or K. The distribution of the ions M(+), O2(+) and M2CrO4(+) in the measured mass spectrum was found to depend on the material used for the Knudsen cell, even for materials such as platinum and gold. In the case of sodium chromate, the decomposition reaction appears to be less important than the molecular vaporization reaction. A preliminary value of 72 kcal/mole at 1141 K was measured for the heat of the molecular vaporization reaction for sodium chromate. In the case of potassium chromate, it has not been possible to conclude which mode of vaporization dominates. For potassium chromate a value of 101 kcal/mole at 1173 K was obtained for the heat of the molecular vaporization reaction.

Stearns, C. A.↗

In Situ Carbon and Sulfur Isotope Analysis of Archean Organic Matter and Pyrite

Stable isotopic compositions of biologically important elements (e.g., C and S) in sedimentary rocks are valuable biosignatures to the extent that they indicate the presence and variable expression of microbial metabolisms in space and time. Strong interactions between the carbon and sulfur cycles (e.g., via organic matter remineralization during microbial sulfate reduction) make coordinated, in situ C and S isotope analysis by secondary ion mass spectrometry (SIMS) a particularly powerful tool. In rocks ranging in age from 2.7-2.5 Ga, expansions in the ranges of delta C-13 of organic matter and delta S-34 of pyrite likely reflect the increasing influence of oxygenic photosynthesis in the surface ocean (as well as methane and sulfur metabolisms in deeper waters), whereas the large range of mass independent sulfur isotope fractionation (Delta S-33) suggests that the atmosphere remained anoxic until approx 2.4 Gyr ago. We report in situ delta C-13 measurements of organic matter in the approx 2.7-2.6 Ga Carawine Dolomite, Marra Mamba Iron Formation, and Jeerinah, Wittenoom, and Tumbiana Formations, as well as the approx 2.5 Ga Mount McRae Shale. We also report in situ delta S-24 and Delta S-33 measurements of pyrite associated with organic matter in a subset of these samples. In a single square cm sample of the Tumbiana Formation with bulk delta C-13(sub org) of -49.7% (VPDB), two distinct kerogen types have delta C-13 values, measured in situ, consistent with oxygenic photosynthesis (-33%) and methane metabolism (-52%). In a sample from the ABDP-9 core, radiobitumen associated with a uraniferous mineral grain is C-13-enriched by 8% (-26.8%0) relative to average in situ kerogen (-34.9%0) and similar in delta C-13 to solvent extractable hydrocarbons from the Mount McRae Shale (avg delta C-13 = -27.1 %). Average reproducibility for delta C-13 was 0.4% (2 SD) using a 6 micron spot and 0.8% using a 3 micron spot. In situ sulfur isotope analyses of 33 authigenic pyrite grains in 3 samples of the ABDP-9 core using a 10 micron spot (2 SD reproducibility = 0.4% for delta S-34 and -0.1% for Delta S-33) show a range of 28.1 % in delta S-34 (-10.3 to 17.8%) and 13.3%0 in Delta S-33 (-3.8 to 9.5%), whereas the range from 132 bulk analyses across 84 m of core is 19.4% for delta S-34 and 11.5% for Delta S-33. Coordinated, in situ carbon and sulfur isotope analyses in one ABDP-9 sample are shown. In situ values from this kerogen-pyrite association are within 0.1% of the bulk value in the case of delta C-13 and higher by several permil in the case of delta S-34 and Delta S-33. Coordinated in situ carbon and sulfur isotope analyses in rocks deposited during key intervals of environmental change (e.g., the Great Oxidation Event) can refine our understanding of the mode and tempo of change. In Earth's oldest sedimentary rocks, and in extraterrestrial samples, these coordinated in situ analyses may reveal biosignatures in the form of isotopic correlations at the scale of individual microorganisms and their microhabitats.

Williford, K. H.↗