Intelligent Integrated Machining: Handheld XRF for Mapping Zr Coating and Al-6061 Cladding Thickness in RERTR Fuel Elements
An inexpensive commercially available handheld energy dispersive x‐ray fluorescence spectrometer is used to provide a rapid non‐destructive estimate of coating and/or cladding thickness in situ . The portable XRF instrument contains a battery‐operated low‐power polychromatic source coupled with a Peltier‐cooled silicon drift detector; built‐in analysis software based upon the Method of Fundamental Parameters converts measured spectral intensities to elemental concentrations, reported as normalized weight percentages. Using detected signals from elements unique to the base material and coating or cladding materials, calibration curves and associated empirical algorithms for determining the thickness of the high‐purity zirconium coating on an RERTR fuel element prior to cladding with aluminum, and for determining the thickness of the Al‐6061 cladding after HIP for a known zirconium layer thickness, are constructed. For Zr over bare DU‐10Mo, agreement between the measured and the calculated thickness is ± 5% or better. For the multilayer system consisting of Al‐6061 over Zr over Du‐10Mo, agreement between calculated and measured thickness is ± 10‐15% over the thickness range of interest. Simultaneous estimates for the Al cladding and underlying Zr coating can be made using the elemental composition output only when the Mo signal is in the quasi‐static range (corresponding to Al thicknesses of ~15 mils (0.381 mm) and greater) and not so low that measurement error is greater than the signal itself; under these conditions, accuracy is on the order of ±15 – 20%. More accurate determination of both Zr and Al thickness simultaneously may be possible if the count‐rates associated with individual fluorescent lines in the multilayer sample are available rather than just the weight percent composition currently output by the XRF analyzer software.