Commissioning and field tests of a van-mounted system for the detection of radioactive sources and Special Nuclear Material

MODES SNM project aimed to carry out technical research in order to develop a prototype for a mobile, modular detection system for radioactive sources and Special Nuclear Materials (SNM). Its main goal was to deliver a tested prototype of a modular mobile system capable of passively detecting weak or shielded radioactive sources with accuracy higher than that of currently available systems. By the end of the project all the objectives have been successfully achieved. Results from the laboratory commissioning and the field tests will be presented. (authors)

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

SU-F-T-368: Improved HPGe Detector Precise Efficiency Calibration with Monte Carlo Simulations and Radioactive Sources

Purpose: To obtain an improved precise gamma efficiency calibration curve of HPGe (High Purity Germanium) detector with a new comprehensive approach. Methods: Both of radioactive sources and Monte Carlo simulation (CYLTRAN) are used to determine HPGe gamma efficiency for energy range of 0–8 MeV. The HPGe is a GMX coaxial 280 cm{sup 3} N-type 70% gamma detector. Using Momentum Achromat Recoil Spectrometer (MARS) at the K500 superconducting cyclotron of Texas A&M University, the radioactive nucleus {sup 24} Al was produced and separated. This nucleus has positron decays followed by gamma transitions up to 8 MeV from {sup 24} Mg excited states which is used to do HPGe efficiency calibration. Results: With {sup 24} Al gamma energy spectrum up to 8MeV, the efficiency for γ ray 7.07 MeV at 4.9 cm distance away from the radioactive source {sup 24} Al was obtained at a value of 0.194(4)%, by carefully considering various factors such as positron annihilation, peak summing effect, beta detector efficiency and internal conversion effect. The Monte Carlo simulation (CYLTRAN) gave a value of 0.189%, which was in agreement with the experimental measurements. Applying to different energy points, then a precise efficiency calibration curve of HPGe detector up to 7.07 MeV at 4.9 cm distance away from the source {sup 24} Al was obtained. Using the same data analysis procedure, the efficiency for the 7.07 MeV gamma ray at 15.1 cm from the source {sup 24} Al was obtained at a value of 0.0387(6)%. MC simulation got a similar value of 0.0395%. This discrepancy led us to assign an uncertainty of 3% to the efficiency at 15.1 cm up to 7.07 MeV. The MC calculations also reproduced the intensity of observed single-and double-escape peaks, providing that the effects of positron annihilation-in-flight were incorporated. Conclusion: The precision improved gamma efficiency calibration curve provides more accurate radiation detection and dose calculation for cancer radiotherapy treatment.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Management of Radioactive Wastes (RWs) and Disused Sealed Radioactive Sources (DSRSs) in Bangladesh: Recent Activities and Challenges Ahead - 16039

The use of radioactive sources in different applications and practices gives rise to various types of RWs that may be detrimental to human health and the environment if not properly managed. Since early sixties of the last century, radioisotopes have found diversified and incremental uses in Bangladesh in the field of medicine, industry, agriculture, food irradiation, research and education and so on. Currently, the country is operating nuclear facilities like 3 MW TRIGA research reactor (TRR), radioisotope production (RIP) facility as well as particle accelerators. Recently, the country has decided to establish nuclear power plants in order to cope with its increasing demand of electric power. The first nuclear power plant is anticipated to come into operation by 2021-2022. Consequently, a notable amount and range of RWs including DSRSs are being expected to be generated from these diversified applications of radioactive materials in near future. Bangladesh has limited facilities and capabilities to manage the RWs which are being currently generated from the conventional activities and practices with radioactive materials. However, the country is also working to improve its necessary capabilities and infrastructures for the management of RWs expected to be generated from both conventional and nuclear activities in future in a comprehensive, integrated and safer manner, ensuring that no undue burden is imposed on present and future generations and on the environment. This paper illustrates the current activities as well as its future plans on the management of RWs and DSRSs in the country. (authors)

12 MANAGEMENT OF RADIOACTIVE AND NON-RADIOACTIVE W↗

Anapole moment of neutrinos and radioactive sources near liquid xenon detectors

We show that placing a radioactive source such as 51 Cr near a liquid xenon detector may allow us to detect the contribution induced by the anapole moment to neutrino-electron scattering in the Standard Model (SM) at the 1 − 2⁢𝜎 level. Although the anapole moment of neutrinos induces a scattering rate with the same spectral shape as the neutral and charged current contributions, exposures of ∼ 60 ton × source run at XENONnT or XLZD may be enough to accumulate sufficient statistics for a detection. We also discuss a simple model where the anapole moment of neutrinos is enhanced or decreased with respect to the SM expectation, further demonstrating how a potential measurement of the anapole moment of neutrinos would allow us to constrain new physics.

dark matter direct detection↗

Radioactive Source Shielding Strategies

This report discusses different options for shielding radioactive sources that will be transported in the Ford Raptor with a stop box during source roundups in Inner Mongolia and aims to determine the best shield option. A packaging and shielding plan based on pre-recovery information is also suggested.

73 NUCLEAR PHYSICS AND RADIATION PHYSICS↗

Evaluating and Countering the Insider Threat to the Radioactive Source Supply Chain

The modern supply chain is a global enterprise and little drove this home more than the global COVID-19 pandemic which sent economic shockwaves throughout the world. Many goods became scarce, as products were delayed, in limited supply, or simply not available. The suddenly diminished supply collided with still high demand and led to greatly increased costs. This was particularly true for the radioactive source supply chain. The pandemic introduced extensive delays for construction projects, slowed the transport of radiological materials to facilities, interrupted treatment deliveries, and impaired the mobility of contractors across the industry. All of these concerns not only adversely affected the economy, but also impacted the safety and security of radiological material, potentially raising national security concerns. The vulnerability of the supply chain, a critical element in an increasingly interconnected world, was exposed. One example that challenged the adaptive capacity of the overall supply chain is the Ever Given container ship, which became stuck in the Suez Canal in 2020. This accident immediately shut down shipments that accounted for 12% of global trade, with long-term impacts estimated to be much larger. Developing the ability to anticipate and react in real-time to sudden changes has quickly become a necessity, particularly in industries that deal with the transport of hazardous material. The reaction to these dramatic incidents was to largely focus attention and resources on protecting the supply chain from external threats. However, the threat to the radioactive material supply chain from insiders intimately involved in the process may be even greater and remains a blind spot that requires increased attention. Recent events revealed the blueprint for targeting and disrupting that supply chain, so the potential for a malicious insider—or a manipulated, unwitting insider—to take advantage of this vulnerability is elevated, creating security concerns for radiological industries. This paper examines and analyzes the potential insider threat to the radioactive source supply chain and recommends steps to take to counter this possibility.

Kinney, Justin↗

Management of Spent and Disused Sealed Radioactive Sources in the Czech Republic - 12124

The Czech Republic is a country with a well-developed peaceful utilization of nuclear energy and ionizing radiation. Sealed Radioactive Sources (further also SRS) are broadly used in many areas in the Czech Republic, e.g. in research, industry, medicine, education, agriculture, etc. Legislation in the field of ionizing radiation source utilization has been fully harmonized with European Community legislation. SRS utilization demands a proper system which must ensure the safe use of SRS, including the management of disused (spent) and orphaned SRS. In the Czech Republic, a comprehensive system of SRS management has been established that is comparable with systems in other developed countries. The system covers both legal and institutional aspects. The Central Register of Ionizing Radiation Sources is an important part of the system. It is a tracking system that covers all activities related to SRS, from their production or import to the end of their use (recycling or disposal). Many spent SRS are recycled and can be used for other purposes after inspection, repacking or reprocessing. When the disused SRS are not intended for further use, they are managed as radioactive waste (RAW). The system of SRS management also ensures the suitable resolution of situations connected with improper SRS handling (in the case of orphaned sources, accidents, etc.). (author)

07 ISOTOPE AND RADIATION SOURCES↗

Inventory and Characterization of Categories 3-5 Disused Sealed Radioactive Sources - 19284

Brazil has one of the biggest inventories of SRS in the world, not only because of the large industrial park, the well-developed health care infrastructure and the numerous mining operations, but also by the use of radioactive sources in consumer products, such as smoke detectors, lightning arresters, surge protectors and even cardiac pacemakers. Presently, the inventory of DSRS at the WML is of more than 10,000; if the sources from lightning arresters, smoke detectors and Ir-192 with activities above the clearance levels are included, the estimate is over 150,000. The Waste Management Laboratory (WML) of the Nuclear and Energy Research Institute (IPEN/CNEN-SP) is responsible for the management and treatment of these sources. An important aspect to consider is that Brazil does not have a defined strategy or a protocol for screening and inventorying these sources, whose management was limited only to storage. The WML started the development of methods and operational work, to characterize and reduce the total volume of this inventory for the borehole-type final disposal facility yet to be constructed, where the DSRS will be disposed of. The information available from these sources is in a database created in 1980, when the sources began to be received. Several records on the FRSD that were checked are incomplete or simply incorrect, since there is no correspondence between the dates of entry of these sources in the storage and their physical characteristics. Some reasons can be listed to try to understand the divergences: the inspection when receiving the sources was not possible, no inspection procedure was defined, there were technical difficulties to open the packages. The objective of this work is to organize the inventory in a standardized way, control access, traceability, reduction of the necessary time to locate and characterize the sources upon reception, and of the repetitive work in the operational routine, and finally release of space in the storage. (authors)

12 MANAGEMENT OF RADIOACTIVE AND NON-RADIOACTIVE W↗

DEPO-related to Radioactive Sources.

Design and Evaluation Process Outline (DEPO) is discussed as it pertains to protection of radioactive sources. The bulk of the report describes features of various kinds of detection systems, and follows this with systems for entry control and personnel identification.

98 NUCLEAR DISARMAMENT, SAFEGUARDS, AND PHYSICAL P↗

The radioactive source calibration system of the PROSPECT reactor antineutrino detector

The Precision Reactor Oscillation and Spectrum (PROSPECT) Experiment is a reactor neutrino experiment created to search for sterile neutrinos with a mass on the order of 1 eV/c 2 and to measure the spectrum of electron antineutrinos from a highly-enriched 235 U nuclear reactor. The PROSPECT detector consists of an 11 by 14 array of optical segments in 6 Li-loaded liquid scintillator at the High Flux Isotope Reactor in Oak Ridge National Laboratory. Antineutrino events are identified via inverse beta decay and read out by photomultiplier tubes located at the ends of each segment. The detector response is characterized using a radioactive source calibration system. Here, we describe the design, operation, and performance of the PROSPECT source calibration system.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

Chapter 9. Control of Accountable Sealed Radioactive Sources (ASRSs) and Radioisotope Thermoelectric Generators (RTGs)

This chapter describes the processes that Members of the Workforce (MOW) follow to control Accountable Sealed Radioactive Sources (ASRSs) and Radioisotope Thermoelectric Generators (RTGs). It also describes the Sandia National Laboratories (SNL) program for implementing the ASRS requirements in 10 CFR 835.1202 and the ASRS and RTG requirements in DOE 0 231.1B.

07 ISOTOPE AND RADIATION SOURCES↗

Innovative methodology for intercomparison of radionuclide calibrators using short half-life in situ prepared radioactive sources

Purpose: An original radionuclide calibrator method for activity determination is presented. The method could be used for intercomparison surveys for short half-life radioactive sources used in Nuclear Medicine, such as{sup 99m}Tc or most positron emission tomography radiopharmaceuticals. Methods: By evaluation of the resulting net optical density (netOD) using a standardized scanning method of irradiated Gafchromic XRQA2 film, a comparison of the netOD measurement with a previously determined calibration curve can be made and the difference between the tested radionuclide calibrator and a radionuclide calibrator used as reference device can be calculated. To estimate the total expected measurement uncertainties, a careful analysis of the methodology, for the case of{sup 99m}Tc, was performed: reproducibility determination, scanning conditions, and possible fadeout effects. Since every factor of the activity measurement procedure can influence the final result, the method also evaluates correct syringe positioning inside the radionuclide calibrator. Results: As an alternative to using a calibrated source sent to the surveyed site, which requires a relatively long half-life of the nuclide, or sending a portable calibrated radionuclide calibrator, the proposed method uses a source preparedin situ. An indirect activity determination is achieved by the irradiation of a radiochromic film using {sup 99m}Tc under strictly controlled conditions, and cumulated activity calculation from the initial activity and total irradiation time. The irradiated Gafchromic film and the irradiator, without the source, can then be sent to a National Metrology Institute for evaluation of the results. Conclusions: The methodology described in this paper showed to have a good potential for accurate (3%) radionuclide calibrators intercomparison studies for{sup 99m}Tc between Nuclear Medicine centers without source transfer and can easily be adapted to other short half-life radionuclides.

07 ISOTOPE AND RADIATION SOURCES↗

A Low-Tech, Low-Budget Storage Solution for High Level Radioactive Sources

The need for safe, secure, and economical storage of radioactive material becomes increasingly important as beneficial uses of radioactive material expand (increases inventory), as political instability rises (increases threat), and as final disposal and treatment facilities are delayed (increases inventory and storage duration). Several vendor-produced storage casks are available for this purpose but are often costly — due to the required design, analyses, and licensing costs. Thus the relatively high costs of currently accepted storage solutions may inhibit substantial improvements in safety and security that might otherwise be achieved. This is particularly true in areas of the world where the economic and/or the regulatory infrastructure may not provide the means and/or the justification for such an expense. This paper considers a relatively low-cost, low-technology radioactive material storage solution. The basic concept consists of a simple shielded storage container that can be fabricated locally using a steel pipe and a corrugated steel culvert as forms enclosing a concrete annulus. Benefits of such a system include 1) a low-tech solution that utilizes materials and skills available virtually anywhere in the world, 2) a readily scalable design that easily adapts to specific needs such as the geometry and radioactivity of the source term material), 3) flexible placement allows for free-standing above-ground or in-ground (i.e., below grade or bermed) installation, 4) the ability for future relocation without direct handling of sources, and 5) a long operational lifetime . ‘Le mieux est l’ennemi du bien’ (translated: The best is the enemy of good) applies to the management of radioactive materials – particularly where the economic and/or regulatory justification for additional investment is lacking. Development of a low-cost alternative that considerably enhances safety and security may lead to a greater overall risk reduction than insisting on solutions that remain economically and/or politically ‘out of reach’.

12 MANAGEMENT OF RADIOACTIVE AND NON-RADIOACTIVE W↗

Securing the metal recycling chain for the steel industry by detecting orphan radioactive sources in scrap metal

Experimental tests are reported for the detection of the heavy metal shielding of orphan sources hidden inside scrap metal by using a recently developed muon tomography system. Shielded sources do not trigger alarm in radiation portal commonly employed at the entrance of steel industry using scrap metal. Future systems integrating radiation portals with muon tomography inspection gates will substantially reduce the possibility of accidental melting of radioactive sources securing the use of recycled metal.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Small-angle Compton Scattering to Determine the Depth of a Radioactive Source in Matter

A gamma-ray peak in a spectrum is often accompanied by a discontinuity in the Compton continuum at the peak. The Compton continuum results from Compton scattering in the detector. The discontinuity at a peak results from small-angle Compton scattering by the gamma rays in matter situated directly between the gamma-ray source and the detector. The magnitude of this discontinuity with respect to the gamma-ray peak is therefore an indicator of the amount of material or shielding between the gamma-ray source and the detector. This small-angle scattering was used to determine the depth of highly-enriched uranium (HEU) solution standards in a concrete floor mockup. The empirical results of the use of this small-angle scattering discontinuity in a concrete floor experiment will be described. A Monte Carlo calculation of the experiment will also be described. In addition, the depth determined from small-angle scattering was used in conjunction with differential attenuation to more accurately measure the uranium content of the mockup. Following these empirical results, the theory of small-angle scattering will be discussed. The magnitude of the discontinuity compared to the peak count rate is directly related to the depth of the gamma-ray source in matter. This relation can be described by relatively simple mathematical expressions. This is the first instance that we are aware of in which the small-angle Compton scattering has been used to determine the depth of a radioactive source. Furthermore this is the first development of the theoretical expressions for the magnitude of the small-angle scattering discontinuity.

73 NUCLEAR PHYSICS AND RADIATION PHYSICS↗

Calibration of a time-resolved hard-x-ray detector using radioactive sources

A four-channel, time-resolved, hard x-ray detector (HXRD) has been operating at the Laboratory for Laser Energetics for more than a decade. The slope temperature of the hot-electron population in direct-drive inertial confinement fusion experiments is inferred by recording the hard x-ray radiation generated in the interaction of the electrons with the target. Measuring the energy deposited by hot electrons requires an absolute calibration of the hard x-ray detector. A novel method to obtain an absolute calibration of the HXRD using single photons from radioactive sources was developed, which uses a thermoelectrically cooled, low-noise, charge-sensitive amplifier.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Radioactive source calibration test of the CMS Hadron Endcap Calorimeter test wedge with Phase I upgrade electronics

The Phase I upgrade of the CMS Hadron Endcap Calorimeters consists of new photodetectors (Silicon Photomultipliers in place of Hybrid Photo-Diodes) and front-end electronics. The upgrade will eliminate the noise and the calibration drift of the Hybrid Photo-Diodes and enable the mitigation of the radiation damage of the scintillators and the wavelength shifting fibers with a larger spectral acceptance of the Silicon Photomultipliers. The upgrade also includes increased longitudinal segmentation of the calorimeter readout, which allows pile-up mitigation and recalibration due to depth-dependent radiation damage. As a realistic operational test, the responses of the Hadron Endcap Calorimeter wedges were calibrated with a $^{60}$Co radioactive source with upgrade electronics. The test successfully established the procedure for future source calibrations of the Hadron Endcap Calorimeters. Lastly we describe the instrumentation details and the operational experiences related to the sourcing test.