Experimental Report: Multi-Instrument Comparison of AAF Size Distribution Instruments
Aerosols are particles suspended in the atmosphere, ranging in size from nanometers to micrometers. Their size distribution affects key atmospheric processes, including nucleation, coagulation, scavenging, activation, and radiative properties (Seinfeld and Pandis 2016). Aerosol size distribution is a critical parameter in atmospheric science, influencing processes such as cloud formation and radiative forcing. Accurate representation of aerosol size distributions is essential for understanding their impact on climate, air quality, and human health. However, aerosol size and composition vary significantly across time and space due to meteorological conditions and natural or anthropogenic sources. (Wu and Boor 2021). Various instruments are used to measure aerosol size distributions, each with distinct principles, advantages, and limitations. This report begins with an in-depth overview of aerosol size-distribution comparison studies, focusing on the passive cavity aerosol spectrometer probe (PCASP), portable optical particle spectrometer (POPS), ultra-high-sensitivity aerosol spectrometer (UHSAS), aerodynamic particle sizer (APS), and scanning mobility particle sizer (SMPS). All of these instruments are used by the U.S. Department of Energy Atmospheric Radiation Measurement (ARM) Aerial Facility (AAF), which commissioned this comparison and report. The report evaluates the strengths and weaknesses of these instruments, highlights their applications, and discusses efforts to merge data from multiple instruments for comprehensive analysis.