Improved thermoelectrically cooled quartz crystal microbalance
Design changes in the thermoelectrically-cooled quartz microbalance, which is used to monitor surface contamination in space simulation chambers, is described in terms of its extended temperature range, increased temperature control, mass sensitivity, and cooling power. The mass sensor uses 20 MHz quartz crystals having a sensitivity of 8.8 x 10 to the minus tenth power g/sq cm - Hz. The crystals are optically polished, metal plated, and overplated with magnesium fluoride to simulate an optical surface. The microbalance temperature circuitry is designed to readout and control surface temperature between 100 C and minus 59 C to plus or minus 0.5 C, and readout only temperature between minus 60 C and minus 199 C using auxiliary liquid nitrogen cooling. Data is included on the measurement of oil contamination of surfaces as a function of temperature in space simulation chambers.