Comparison of UNL laser imaging and sizing system and a phase/Doppler system for analyzing sprays from a NASA nozzle
Aerosol spray characterization was done using a P/DPA and a laser imaging/video processing system on a NASA MOD-1 air-assist nozzle being evaluated for use in aircraft icing research. Benchmark tests were performed on monodispersed particles and on the NASA MOD-1 nozzle under identical laboratory operating conditions. The laser imaging/video processing system and the P/DPA showed agreement on calibration tests in monodispersed aerosol sprays of + or - 2.6 microns with a standard deviation of + or - 2.6 microns. Tests were performed on the NASA MOD-1 nozzle on the centerline and radially at one-half inch increments to the outer edge of the spray plume at a distance two feet (0.61 m) downstream from the exit of the nozzle. Comparative results at two operating conditions of the nozzle are presented for the two instruments. For the first case, the deviation in arithmetic mean diameters determined by the two instruments was in a range of 0.1 to 2.8 microns, and the deviation in Sauter mean diameters varied from 0 to 2.2 microns. Operating conditions in the second case were more severe which resulted in the arithmetic mean diameter deviating from 1.4 to 7.1 microns and the deviation in the Sauter mean diameters ranging from 0.4 to 6.7 microns.