Large aperture interferometer with phase-conjugate self-reference beam

A large aperture self-referencing interferometer consisting of a Twyman-Green interferometer using a self-pumped phase conjugator in series with test section optics is described and experimentally demonstrated. This interferometer provides twice the fringe shift of a Mach-Zehnder (M-Z) interferometer for a given optical phase change induced within the test section. It also provides greater irradiance in the reference beam than does a similar series setup utilizing a M-Z interferometer incorporating a local reference beam. Whereas the ordinary interferometer records instantaneous conditions, the new one records overage conditions if a BaTiO3 crystal is used as the phase conjugator.

Howes, W. L.↗

The Mark III stellar interferometer

The Mark III interferometer is an operational long baseline stellar interferometer on Mt. Wilson with four-possible baseline configurations from 9m NE-SW to 20m N-S. The interferometer was designed to be a highly automated astronomical instrument to measure stellar positions and diameters to a magnitude limit of seven. Initial fringe observations were made in September 1986 with a 12-m N-S baseline. In the following months, semi-automated astrometric and stellar diameter measurements were also made. This paper describes the hardware and software components of the instrument and its operational characteristics. The interferometer has several novel features. One is the use of optimal estimation and control algorithms (e.g. Kalman filters) in the control loops. Another is the ability to operate both as a closed-loop phased interferometer and eventually as an open-loop or absolute coherent interferometer. High thermal stability and mechanical accuracy should permit the instrument to point blind at an astronomical object and maintain optical path equality to within the limits set by the atmosphere. In this absolute interferometric mode of operation, it should be possible to observe faint astronomical objects that are too dim for phase tracking. In theory, measurements of amplitude, group delay, and closure phase will be possible to 14 mag.

Shao, M.↗

Phase conjugate Twyman-Green interferometer for testing spherical surfaces and lenses and for measuring refractive indices of liquids or solid transparent materials

The present paper describes an application of a phase conjugate Twyman-Green interferometer using barium titanate as a self-pumping mirror for testing optical components like concave and convex spherical mirrors and lenses. The aberrations introduced by the beam splitter while testing concave or convex spherical mirrors of large aperture are automatically eliminated due to self-focussing property of the phase conjugate mirror. There is no necessity for a good spherical surface as a reference surface unlike in classical Twyman-Green interferometer or Williams interferometer. The phase conjugate Twyman Green interferometer with a divergent illumination can be used as a test plate for checking spherical surfaces. A nondestructive technique for measuring the refractive indices of a Fabry Perot etalon by using a phase conjugate interferometer is also suggested. The interferometer is found to be useful for measuring the refractive indices of liquids and solid transparent materials with an accuracy of the order of + or - 0.0004.

Shukla, R. P.↗

Interferometer-Controlled Optical Tweezers Constructed for Nanotechnology and Biotechnology

A new method to control microparticles was developed in-house at the NASA Glenn Research Center in support of the nanotechnology project under NASA's Aerospace Propulsion and Power Base Research Program. A prototype interferometer-controlled optical tweezers was constructed to manipulate scanning probe microscope (SPM) tips. A laser beam passed through a Mach-Zehnder interferometer, and a microscope objective then produced an optical trap from the coaxial beams. The trap levitated and generated the coarse motion of a 10-mm polystyrene sphere used to simulate a SPM tip. The interference between the beams provided fine control of the forces and moments on the sphere. The interferometer included a piezoelectric-scanned mirror to modulate the interference pattern. The 10-mm sphere was observed to oscillate about 1 mm as the mirror and fringe pattern oscillated. The prototype tweezers proved the feasibility of constructing a more sophisticated interferometer tweezers to hold and manipulate SPM tips. The SPM tips are intended to interrogate and manipulate nanostructures. A more powerful laser will be used to generate multiple traps to hold nanostructures and SPM tips. The vibrating mirror in the interferometer will be replaced with a spatial light modulator. The modulator will allow the optical phase distribution in one leg of the interferometer to be programmed independently at 640 by 480 points for detailed control of the forces and moments. The interference patterns will be monitored to measure the motion of the SPM tips. Neuralnetwork technology will provide fast analysis of the interference patterns for diagnostic purposes and for local or remote feedback control of the tips. This effort also requires theoretical and modeling support in the form of scattering calculations for twin coherent beams from nonspherical particles.

Decker, Arthur J.↗

Solar CIV Vacuum-Ultraviolet Fabry-Perot Interferometers

Aims: A tunable, high spectral resolution, high effective finesse, vacuum ultraviolet (VUV) Fabry-Perot interferometer (PPI) is designed for obtaining narrow-passband images, magnetograms, and Dopplergrams of the transition region emission line of CIV (155 nm). Methods: The integral part of the CIV narrow passband filter package (with a 2-10 pm FWHM) consists of a multiple etalon system composed of a tunable interferometer that provides high-spectral resolution and a static low-spectral resolution interferometer that allows a large effective free spectral range. The prefilter for the interferometers is provided by a set of four mirrors with dielectric high-reflective coatings. A tunable interferometer, a VUV piezoelectric-control etalon, has undergone testing using the surrogate F2 eximer laser line at 157 nm for the CIV line. We present the results of the tests with a description of the overall concept for a complete narrow-band CIV spectral filter. The static interferometer of the filter is envisioned as being hudt using a set of fixed MgF2 plates. The four-mirror prefilter is designed to have dielectric multilayer n-stacks employing the design concept used in the Ultraviolet Imager of NASA's Polar Spacecraft. A dual etalon system allows the effective free spectral range to be commensurate with the prefilter profile. With an additional etalon, a triple etalon system would allow a spectrographic resolution of 2 pm. The basic strategy has been to combine the expertise of spaceflight etalon manufacturing with VUV coating technology to build a VUV FPI which combines the best attributes of imagers and spectrographs into a single compact instrument. Results. Spectro-polarimetry observations of the transition region CIV emission can be performed to increase the understanding of the magnetic forces, mass motion, evolution, and energy release within the solar atmosphere at the base of the corona where most of the magnetic field is approximately force-free. The 2D imaging of the full vector magnetic field at the height of maximum magnetic influence (minimum plasma beta) can be accomplished, albeit difficult, by measuring the Zeeman splitting of the CIV resonance pair. Designs of multiple VUV FPIs can be developed for integration into future orbiting solar observatories to obtain rapid cadence, spectral imaging of the transition region.

Gary, G. Allen↗

Parallel Wavefront Analysis for a 4D Interferometer

This software provides a programming interface for automating data collection with a PhaseCam interferometer from 4D Technology, and distributing the image-processing algorithm across a cluster of general-purpose computers. Multiple instances of 4Sight (4D Technology s proprietary software) run on a networked cluster of computers. Each connects to a single server (the controller) and waits for instructions. The controller directs the interferometer to several images, then assigns each image to a different computer for processing. When the image processing is finished, the server directs one of the computers to collate and combine the processed images, saving the resulting measurement in a file on a disk. The available software captures approximately 100 images and analyzes them immediately. This software separates the capture and analysis processes, so that analysis can be done at a different time and faster by running the algorithm in parallel across several processors. The PhaseCam family of interferometers can measure an optical system in milliseconds, but it takes many seconds to process the data so that it is usable. In characterizing an adaptive optics system, like the next generation of astronomical observatories, thousands of measurements are required, and the processing time quickly becomes excessive. A programming interface distributes data processing for a PhaseCam interferometer across a Windows computing cluster. A scriptable controller program coordinates data acquisition from the interferometer, storage on networked hard disks, and parallel processing. Idle time of the interferometer is minimized. This architecture is implemented in Python and JavaScript, and may be altered to fit a customer s needs.

Rao, Shanti R.↗

Quantum noise of a Michelson-Sagnac interferometer with a translucent mechanical oscillator

Quantum fluctuations in the radiation pressure of light can excite stochastic motions of mechanical oscillators thereby realizing a linear quantum opto-mechanical coupling. When performing a precise measurement of the position of an oscillator, this coupling results in quantum radiation pressure noise. Up to now this effect has not been observed yet. Generally speaking, the strength of radiation pressure noise increases when the effective mass of the oscillator is decreased or when the power of the reflected light is increased. Recently, extremely light SiN membranes ({approx_equal}100 ng) with high mechanical Q values at room temperature ({>=}10{sup 6}) have attracted attention as low thermal noise mechanical oscillators. However, the power reflectance of these membranes is much lower than unity (<0.4 at a wavelength of 1064 nm) which makes the use of advanced interferometer recycling techniques to amplify the radiation pressure noise in a standard Michelson interferometer inefficient. Here, we propose and theoretically analyze a Michelson-Sagnac interferometer that includes the membrane as a common end mirror for the Michelson interferometer part. In this topology, both power and signal recycling can be used even if the reflectance of the membrane is much lower than unity. In particular, signal recycling is a useful tool because it does not involve a power increase at the membrane. We derive the formulas for the quantum radiation pressure noise and the shot noise of an oscillator position measurement and compare them with theoretical models of the thermal noise of a SiN membrane with a fundamental resonant frequency of 75 kHz and an effective mass of125 ng. We find that quantum radiation pressure noise should be observable with a power of 1 W at the central beam splitter of the interferometer and a membrane temperature of 1 K.

71 CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSIC↗

Four-point bisensitivity velocity interferometer with a multireflection etalon

A four-point bisensitivity velocity interferometer system for any reflector (VISAR) with a renovative delay etalon is proposed and demonstrated. In this interferometer, we introduce a new film-coating strategy to accurately measure small velocity with relatively short and cheap etalon. Laser pointing to the etalon is split into two beams with different incident angles with each beam going through the etalon in different path. The beam with the smaller incident angle is reflected three times before it leaves the etalon, while the other beam with larger incident angle goes through the etalon to and forth only once. The delay time of the laser beam with smaller incident angle is almost three times longer than that of the beam with larger incident angle. In the example of the laser with a smaller incident angle, the velocity per fringe of this interferometer can be reduced by approximately three times. The etalon is optimized so that four laser beams can be penetrated in the vertical direction at the meantime. With an etalon of 200 mm in diameter and 150 mm in length, a four-point bisensitivity velocity interferometer can achieve the velocity per fringe of 100 and 350 m/s fringe. A measurement has been successfully undertaken for the steel flyer driven by the explosive where the developed interferometer applies.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Admissible Crosstalk Limits in a Two Colour Interferometers for Plasma Density Diagnostics. A Reduction Algorithm

Mid Infrared (10,6 {mu}m CO{sub 2} laser lines) interferometers as a plasma density diagnostic must use two-colour systems with superposed interferometers beams at different wavelengths in order to cope with mechanical vibrations and drifts. They require a highly precise phase difference measurement where all sources of error must be reduced. One of these is the cross-talk between the signals which creates nonlinear spurious periodic mixing products. The reason may be either optical or electrical crosstalk both resulting in similar perturbations of the measurement. In the TJII interferometer a post-processing algorithm is used to reduce the crosstalk in the data. This post-processing procedure is not appropriate for very long pulses, as it is the case for in new tokamak (ITER) or stellarator (W7-X) projects. In both cases an on-line reduction process is required or--even better--the unwanted signal components must be reduced in the system itself CO{sub 2} laser interferometers which as the second wavelength use the CO laser line (5,3 {mu}m), may apply a single common detector sensitive to both wavelengths and separate the corresponding IF signals by appropriate bandpass filters. This reduces complexity of the optical arrangement and avoids a possible source of vibration induced phase noise as both signals share the same beam path. To avoid cross talk in this arrangement filtering must be appropriate. In this paper we present calculations to define the limits of crosstalk for a desired plasma density precision. A crosstalk reduction algorithm has been developed and is applied to experimental results from TJ-II pulses. Results from a single detector arrangement as under investigation for the CO{sub 2}/CO laser interferometer developed for W7-X are presented.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

A generalized, periodic nonlinearity-reduced interferometer for straightness measurements

Periodic nonlinearity is a systematic error limiting the accuracy of displacement measurements at the nanometer level. However, an interferometer with a displacement measurement accuracy of less than 1 nm is required in nanometrology and in fundamental scientific research. To meet this requirement, a generalized, periodic nonlinearity-reduced interferometer, based on three construction principles has been developed for straightness measurements. These three construction principles have resulted in an interferometer with a highly stable design with reduced periodic nonlinearity. Verifications by a straightness interferometer have demonstrated that the periodic nonlinearity was less than 40 pm. The results also demonstrate that the interferometer design is capable of subnanometer accuracy and is useful in nanometrology.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Multichannel microwave interferometer for the levitated dipole experiment

A four-channel microwave interferometer (center frequency: 60 GHz) has been constructed to measure plasma density profiles in the levitated dipole experiment (LDX). The LDX interferometer has a unique design owing to the unique geometry of LDX. The main design features of the interferometer are: (1) the transmitted beam traverses the plasma entirely in O-mode; (2) the interferometer is a heterodyne system employing two free-running oscillators; (3) four signals of data are received from just on transmitted beam; (4) phase shifts are detected in quadrature. Calibration tests demonstrate that the interferometer measures phase shifts with an uncertainty of approximately 5 deg. Plasma densities in LDX corresponding to phase shifts of up to 5{pi} are routinely and successfully measured.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

X-ray Talbot-Lau interferometer for high-speed phase imaging and tomography using white synchrotron radiation

The Talbot interferometer using white synchrotron radiation has already been demonstrated for high-speed X-ray phase imaging and tomography as well as four-dimensional phase tomography for an observation of a dynamic specimen. In those previous experiments, the grating lines were oriented horizontally because the synchrotron radiation source size is large in the horizontal direction, and only the vertical spatial coherence satisfies the requirement for the operation of the Talbot interferometer given its distance from the source. For non-rigid samples, the horizontal axis of rotation causes unwanted motion of the sample due to gravity which results to artifact in the tomography reconstruction. For fluid samples, a vertical rotation axis is certainly necessary. While it is possible to orient the sample rotation axis perpendicular to the grating lines of the Talbot interferometer, solving the definite integral of the differential phase images to obtain the phase shift for x-ray phase tomography proves to be cumbersome when the sample extends outside the image and there is no null region for which the integration constant is known. In this work we aimed at increasing the spatial coherence of the x-rays along the horizontal so that the grating lines and the sample rotation axis could be oriented vertically. A Talbot-Lau interferometer was constructed by adding an absorption grating which acted as vertical line sources of horizontally spatially coherent white synchrotron radiation to the Talbot interferometer. An average of 20% moire fringe visibility was obtained. The set-up was demonstrated for highspeed phase tomography of a polymer sample.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Dispersion interferometer using modulation amplitudes on LHD (invited)

Since a dispersion interferometer is insensitive to mechanical vibrations, a vibration compensation system is not necessary. The CO{sub 2} laser dispersion interferometer with phase modulations on the Large Helical Device utilizes the new phase extraction method which uses modulation amplitudes and can improve a disadvantage of the original dispersion interferometer: measurement errors caused by variations of detected intensities. The phase variation within ±2 × 10{sup 17} m{sup −3} is obtained without vibration compensation system. The measured line averaged electron density with the dispersion interferometer shows good agreement with that with the existing far infrared laser interferometer. Fringe jump errors in high density ranging up to 1.5 × 10{sup 20} m{sup −3} can be overcome by a sufficient sampling rate of about 100 kHz.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Orientational atom interferometers sensitive to gravitational waves

We present an atom interferometer that differs from common atom interferometers as it is not based on the spatial splitting of electronic wave functions, but on orienting atoms in space. As an example we present how an orientational atom interferometer based on highly charged hydrogen-like atoms is affected by gravitational waves. We show that a monochromatic gravitational wave will cause a frequency shift that scales with the binding energy of the system rather than with its physical dimension. For a gravitational wave amplitude of h=10{sup -23} the frequency shift is of the order of 110 {mu}Hz for an atom interferometer based on a 91-fold charged uranium ion. A frequency difference of this size can be resolved by current atom interferometers in 1 s.

71 CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSIC↗

102({h_bar}/2{pi})k Large Area Atom Interferometers

We demonstrate atom interferometers utilizing a novel beam splitter based on sequential multiphoton Bragg diffractions. With this sequential Bragg large momentum transfer (SB-LMT) beam splitter, we achieve high contrast atom interferometers with momentum splittings of up to 102 photon recoil momenta (102({h_bar}/2{pi})k). To our knowledge, this is the highest momentum splitting achieved in any atom interferometer, advancing the state-of-the-art by an order of magnitude. We also demonstrate strong noise correlation between two simultaneous SB-LMT interferometers, which alleviates the need for ultralow noise lasers and ultrastable inertial environments in some future applications. Our method is intrinsically scalable and can be used to dramatically increase the sensitivity of atom interferometers in a wide range of applications, including inertial sensing, measuring the fine structure constant, and detecting gravitational waves.

71 CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSIC↗

Measurement of angle and axis of rotation in a carousel interferometer: a detailed analysis

A detailed analysis of a carousel interferometer is presented for the measurement of an unknown angle and axis of rotation. The technique exploits a set of compensator glass plates and a right-angle prism that is placed in each of the two arms of the interferometer. The two sets are placed at the same rotational stage, while the end mirrors of the interferometer are static. When rotation takes place, individual and relative optical path differences are generated in the two beams of the interferometer. The generated phase differences contribute toward finding the angle and axis of rotation. The analysis is presented for any initial position of the interferometer, i.e., the radial vector from the axis of rotation to the apex of one of the prisms used. The results show the slight variations in the error and nonlinearity when different parameters are manipulated. Moreover, the trade-off between the maximum size of the prisms and the radial distances are also presented.

71 CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSIC↗

Strongly coupled edge states in a graphene quantum Hall interferometer

Electronic interferometers using the chiral, one-dimensional (1D) edge channels of the quantum Hall effect (QHE) can demonstrate a wealth of fundamental phenomena. The recent observation of phase jumps in a Fabry-Pérot (FP) interferometer revealed anyonic quasiparticle exchange statistics in the fractional QHE. When multiple integer edge channels are involved, FP interferometers have exhibited anomalous Aharonov-Bohm (AB) interference frequency doubling, suggesting putative pairing of electrons into 2e quasiparticles. Here, we use a highly tunable graphene-based QHE FP interferometer to observe the connection between interference phase jumps and AB frequency doubling, unveiling how strong repulsive interaction between edge channels leads to the apparent pairing phenomena. By tuning electron density in-situ from filling factor v < 2 to v > 7, we tune the interaction strength and observe periodic interference phase jumps leading to AB frequency doubling. Our observations demonstrate that the combination of repulsive interaction between the spin-split v = 2 edge channels and charge quantization is sufficient to explain the frequency doubling, through a near-perfect charge screening between the localized and extended edge channels. Our results show that interferometers are sensitive probes of microscopic interactions and enable future experiments studying correlated electrons in 1D channels using density-tunable graphene.

42 ENGINEERING↗

Increased interference fringe visibility from the post-fabrication heat treatment of a perfect crystal silicon neutron interferometer

We find that annealing a previously chemically etched interferometer at 800 °C dramatically increased the interference fringe visibility from 23% to 90%. The Bragg plane misalignments were also measured before and after annealing using neutron rocking curves, showing that Bragg plane alignment was improved across the interferometer after annealing. Furthermore, this suggests that current interferometers with low fringe visibility may be salvageable and that annealing may become an important step in the fabrication process of future neutron interferometers, leading to less need for chemical etching and larger more exotic neutron interferometers.

47 OTHER INSTRUMENTATION↗