Electrical I-V Probes Calibration And Reactor Plasma Chamber Characterization
We have constructed a high resolution I-V Probes including a high voltage and current probes to measure the current and the tension at the entry of the electrode excitation in reactor plasma RF argon. The I-V Probes is no intrusive, therefore it can be transposable in another surface treatment reactor, and it is what is required in the industrial applications. Because of capacitive aspect of the coupling and the presence of parasitic electrical elements in the chamber plasma reactor and the connection circuits, the measured signals are largely different from the signals at the RF electrode in contact with plasma. A phase angle error of 1 deg. between i(t) and v(t) introduce a large error in the calculated parameters. The objective of this work is the calibration of the I-V Probes and the whole system of measurement, and the characterization of the parasitic elements of the reactor. The characterization of the I-V Probes required the realization of several reactive loads (capacitive) measured with a network analyzer HP8753ES. The signals Vi(t) and Vv(t), resulting from the I-V Probes and collected by a DPO (TDS4054 numerical oscilloscope of very high performance), are transferred towards the computer for a digital processing. The elimination of the offsets voltage present at the entries of the DPO and the smoothing by a powerful method of least squares, make it possible to increase precision.